MDO3000 Programmer Manual.pdf - 第971页

Commands Listed in Alphabetical Order Conditions MDO3000 series . Requires option SA3 or SA6 for the MDO4000C series. Requires the DPO4VID application m odule for use with MSO/DPO4000B, MDO4000 and MDO4 000B models. Grou…

100%1 / 1158
Commands Listed in Alphabetical Order
Group
Video Picture
Syntax
VIDPic:LOCation:STARt:LINE <NR1>
VIDPic:LOCation:STARt:LINE?
Arguments
Integer that represents the video picture starting line number, in the range of
0 to 1023. The default is 17.
Examples
VIDPIC:STAR:LINE 348 sets the video picture starting line number to 348.
VIDPic:LOCation:STARt:PIXel
Sets (or queries) the video picture starting pixel in each line
The range for this value varies with the instrument screen geometry. Each video
line is an analog signal of nominally 63.5us duration (10.9us blanking + 52.6us
active video) that can be sampled from 1MS/s to 100MS/s, giving from 52.6 to
5260 video samples or "pixels" per output line. The start pixel is the starting
sample in each line; it is limited to a single line at the current sample rate,i.e.
ranges between -10.9 to 52.6 and -1090 to +5260.
Conditions
MDO3000 series. Requires option SA3 or SA6 for the MDO4000C series.
Requires the DPO4VID application module for use with MSO/DPO4000B,
MDO4000 and MDO4000B models.
Group
Video Picture
Syntax
VIDPic:LOCation:STARt:PIXel <NR1>
VIDPic:LOCation:STARt:PIXel?
Arguments
Integer that represents the video picture starting pixel number. The default is 0.
Examples
VIDP:LOC:STAR:PIX 347 sets the video picture starting pixel number to 347.
VIDPic:LOCation:WIDth
Sets (or queries) the video picture width, in columns.
2-940 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Conditions
MDO3000 series
. Requires option SA3 or SA6 for the MDO4000C series.
Requires the DPO4VID application module for use with MSO/DPO4000B,
MDO4000 and MDO4000B models.
Group
Video Picture
Syntax
VIDPic:LOCation:WIDth <NR1>
VIDPic:LOCation:WIDth?
Arguments
Integer that represents the video picture width, in columns. This value must be >=
1. The range for this value is 1 750. The default is 526.
Examples
VIDPIC:LOC:WID 348 sets the video picture width to 348 columns.
VIDPic:LOCation:X
Sets (or queries) the video picture X origin location, in columns.
Conditions
MDO3000 series. Requires option SA3 or SA6 for the MDO4000C series.
Requires t
he DPO4VID application module for use with MSO/DPO4000B,
MDO4000 and MDO4000B models.
Group
Video Picture
Syntax
VIDPic:LOCation:X <NR1>
VIDPic:LOCation:X?
Arguments
Integer that represents the video picture X origin location, in columns, where
column 0 is the left-most edge of the display. The range for this value is 0 -749.
The default is 112 [ = (750 - 526) / 2 ].
Examples
VIDP:LOC:X 367 sets the video picture X origin location to column 367.
VIDPic:LOCation:Y
Sets (or queries) the video picture Y origin location, in rows.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-941
Commands Listed in Alphabetical Order
Conditions
MDO3000 series
. Requires option SA3 or SA6 for the MDO4000C series.
Requires the DPO4VID application module for use with MSO/DPO4000B,
MDO4000 and MDO4000B models.
Group
Video Picture
Syntax
VIDPic:LOCation:Y <NR1>
VIDPic:LOCation:Y?
Arguments
Integer that represents the video picture Y origin location, in rows, whererow0is
the top most edge of the display. The range for this value is 0 - 399. The default is
73. [ = (400 - 254) / 2 ].
Examples
VIDP:LOC:Y 367 sets the video picture Y origin location to row 367.
VIDPic:SOUrce
Sets (or queries) the channel to use for the video picture source waveform.
Conditions
MDO3000 series. Requires option SA3 or SA6 for the MDO4000C series.
Requires the DPO4VID application module for use with MSO/DPO4000B,
MDO4000 and MDO4000B models.
Group
Video Picture
Syntax
VIDPic:SOUrce {CH1|CH2 |CH3|CH4}
VIDPic:SOUrce?
Arguments
CH1|CH2|CH3|CH4 specify which channel to use for the video picture source
waveform.
Examples
VIDP:SOU CH3 specifies to use Channel 3 as the source waveform.
VIDPic:STANdard
Sets (or queries) which video picture standard to use (either NTSC or PAL).
2-942 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual