Do the same as with the X axis and carry out adjustment of the measured position. Once the reference is determined, select "OK". Selecting "OK" causes measurement of the distance from OCC to the Y dir…

100%1 / 257
Once reference is set, select "OK" and OCC will move to X axis scaling position.
Monitor image when adjustment completed
With teaching adjust the monitor cross hair cursor so that it falls near the center of the 2 marks and
press HOD Enter key.
If no adjustment is needed, teaching is not required.
Should any dust be visible at this time, wipe the camera adjust jig thoroughly so that only the marks
are visible.
Anything visible besides the marks at time of measurement may make it impossible to take a
correct reading or may cause error.
Once the reference is determined, select "OK".
Selecting "OK" causes measurement of the distance from OCC to the X directional mark.
When measurement is completed, the measured value is displayed. From the input value and
measured value, X directional scaling is calculated and modified.
Once the reference is determined, select "OK".
Select "OK" and the OCC will move to the Y axis scaling position.
Monitor image when adjustment completed
2-31
Do the same as with the X axis and carry out adjustment of the measured position.
Once the reference is determined, select "OK".
Selecting "OK" causes measurement of the distance from OCC to the Y directional mark.
When measuring is completed, the measured value is displayed. From the input value and
measured value, Y directional scaling is calculated and modified.
(2) Assembling Angle
Once the reference is determined, select "OK".
Selecting "OK" causes OCC to move to assembling angle measurement position.
Monitor image when adjustment completed
With teaching, adjust the monitor cross hair cursor so that it falls near the center of the mark and
press HOD Enter key.
Anything visible besides the marks at time of measurement will make it impossible to take a correct
reading or may cause error.
Once the reference is determined, select "OK".
Selecting "OK" causes measurement of the assembling angle.
2-32
When measuring is completed, the assembling angle is calculated from the recognition result and
modified.
Remove camera adjustment stage from the CAL block.
Once completed, select "OK".
Setting is completed.
Select "OK" to return to default operation screen.