Service Ma nual W PC5 / WPC6 4.3.3 Calibrating the Limit Switch The plus and m inus position of the limi t switches need to b e calibrated. Calibrating the limit sw itch - ➢ Select the l oad axis (1). ➢ Select the func t…

Service Manual WPC5 / WPC6
➢ Confirm the new reference point proximity switch with (1).
Setting the Driver Cam
➢ Loosen the screw (1), fastening the cam to the driver.
➢ Correct the position of the cam by moving it accordingly in either the – or + direction.
➢ Retighten the screw (1) and repeat the measuring steps (see "3.7.13 Replacing the
Reference Point.
Proximity Switch for the Load Axis [03056947-xx]" [➙ 3-99]).
➢ Wiederholen Sie die einzelnen Schritte, bis keine Fehlermeldung mehr erscheint.

Service Manual WPC5 / WPC6
4.3.3 Calibrating the Limit Switch
The plus and minus position of the limit switches need to be calibrated.
Calibrating the limit switch -
➢ Select the load axis (1).
➢ Select the function (2).
➢ Select (3).
➢ Follow the instruction shown and continue with (1).

Service Manual WPC5 / WPC6
➢ Confirm the new position at which the limit switch triggers with (1).
Calibrating the limit switch +
➢ Select the lifting axis (1).
➢ Select the function and continue with the same procedure as that
used for calibrating the limit switch.