4.3 Calibrating the Load Axis (WPC6 only)
Before calibration, remove all waffle-pack tray carriers (WPTCs) from the WPC.
Calibration of the load axis includes the following functions:
"4.3.2 Teaching the Zero Point Position" [
4-133]
"4.3.3 Reference Proximity Switch (Bero)" [
4-135]
"4.3.4 Calibrating the Limit Switch" [
4-137]
Load axis order of settings
If the limit switches or sensors are replaced, you will need to recalibrate or reset these
accordingly.
A complete recalibration is only needed if the zero point position is changed.
The function for the
limit switch +
and
limit switch
specified in the software is assumed by the
end position stoppers (bumpers) at the machine end and tower end.
Service Manual WPC5 / WPC6
Page 4-133
4.3.1 Teaching the Zero Point Position
Select the load axis (1).
Select the function
Teach zero position
(2).
Select
Next
(3).
Follow the instruction shown and continue with
Commit
(1).