DMM4020.pdf - 第4页
Datasheet Characteristics General Speci fi cations Vo l t a g e Characteristic Description 100V Setting 90 V to 1 10 V 120V Setting 108 V to 132 V 220V Setting 198 V to 242 V 240V Setting 2 1 6Vt o2 6 4V Frequency 47 Hz t…

Digital Multimeters — Tektronix DMM4020
SignalExpress acquiring data from Tektronix DMM4020 and DPO3052.
Connect Your Bench for Intelligent Debug
Easily capture, save, and analyze measurement results from your
multimeter with the special Tektronix Edition of National Instrument’s
LabVIEW S ignalExpress™ software. Every DMM4020 multimeter ships
with a free copy of the Limited Edition version of SignalExpress for basic
instrument cont rol, data logging, and analysis. The optional Professional
Edition offers over 20 0 built-in functions that provide additional signal
processing, advanced analysis, sweeping, limit testing, and user-defined
step capabilities.
SignalExpress support s the range of Tektronix bench instruments*
1
,
enabling you to connect your entire test bench. You can then access the
SignalEx
press used to export DMM4020 data into Excel.
feature-rich tools packed into each instrument from one intuitive software
interface. This allows you to automate complex measurements requiring
multiple instruments, log data for an extended period of time, time correlate
data from multiple instruments, and easily capture and analyze your results,
all from
your PC. Only Tektronix offers a connected test be nch of intelligent
instruments to simplify and speed debug of your complex design.
Performance You Can Count On
In addition to industry-leading service and support, every DMM4020
multimeter comes backed with a three-year standard warranty.
*
1
NI LabVIEW SignalExpress supports the following Tektronix instruments: MSO/DPO4000/3000/2000
Series oscilloscopes, TDS3000C/2000B/1000B Series oscilloscopes, AFG3000 Series arbitrary/function
generators, DMM4050/4040/4020 Series digital multimeters.
www.tektronix.comdmm 3

Datasheet
Characteristics
General Specifications
Voltage
Characteristic
Description
100V Setting
90 V to 110 V
120V Setting
108 V to 132 V
220V Setting
198 V to 242 V
240V Setting
216Vto264V
Frequency 47 Hz to 440 Hz
Power Consumption 15 VA peak (10 W average)
Dimensions
Dimension
mm
in.
Height 88 3.46
Width 217 8.56
Depth 297 11.7
Weight kg lb.
Net 2.1 4.6
Display
Vacuum Fluorescent Display, segment
Environment
Characteristic
Description
Temperature
Operating 0 °C to 50 °C
Storage –40 °C to 70 °C
Warm Up
½ hour to full uncertainty specifications
Relative Humidity (noncondensing)
Operating Uncontrolled (<10 °C)
<90%(10°Cto28°C)
<75%(28°Cto40°C)
<45%(40°Cto50°C)
Storage <95% (–40 °C to 70 °C)
Altitude
Operating
2,000 meters
Storage
12,000 meters
Vibration
Complies with MIL-PRF-28800F Class 3
Safety Complies with IEC 61010-1:2001, ANSI/ISA 61010-1
(S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2
No. 61010.1:2004, CAT I 1000V / CAT II 600 V
EMC Des
ignedtocomplywithIEC
613
26-1:1997+A1:1998+A2:2000
Triggering
Characteristic
Description
Trigger Delay 400 ms
External Trigger Delay <2 ms
External Trigger Jitter <1 ms
Trigger Input TTL Levels
Trigger Output
5Vmax
Math Functio
ns
Min/Max, relative, hold, compare, and dB functions
Multiple Measurement Combinations
Primary Fu nction
Secondary
Function
DC V AC V DC I AC I FREQ OHMS
DC V
XXXX
AC V
XXXXX
DC I
XXXX
AC I
XXXX
FREQ
XX
OHMS
X
Electrical
Characteristic
Description
Input Protection 1000 V all ranges
Overrange 10% on the largest ranges of all functions except
continuity and diode test
Remote Interfaces
RS-232
C (RS-232 to USB Adapter cable included)
Warranty
Three years
4 www.tektronix.comdmm

Digital Multimeters — Tektronix DMM4020
Electrical Specifications
DC Voltage Specifications
Specifications are valid for 5½ digit mode and after at least a half-hour warm up.
Characteristic
Description
Maximum Input 1000 V on any range
Common Mode
Rejection
120 dB at 50 or 60 Hz 0.1% (1 kΩ unbalance)
Normal Mode Rejection
80 dB at Slow Rate
A/D Nonlinearity 15 ppm of range
Input Bias Current <30 pA at 25 °C
Settling Considerations Measurement settling times are affected by source
impedance, cable dielectric characteristics, and input
signal changes
Input Cha
racteristics
ResolutionRange
Full Scale
(5½ Digits)
Slow
Medium Fast
Input Impe dan ce
200 mV 199.999 mV 1 μV10μV10μV
>10 GΩ*
2
2 V 1.99999 V 10 μV100μV100μV
>10 GΩ*
2
20 V 19.9999 V 100 μV1000μV 1000 μV
10 MΩ ±1%
200 V 199.999 V 1 mV 10 mV 10 mV
10 MΩ ±1%
1000 V 1000.00 V 10 mV 100 mV 100 mV
10 MΩ ±1%
*
2
At som
e dual-display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 MΩ.
Accuracy
Uncertainty*
3
90 days 1 year
Range
23 °C ±5 °C 23 °C ±5 °C
Te mp eratu re Coefficient/°C
Outside18-28°C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005
2 V 0.01 + 0.002 0.015 + 0.003 0.001 + 0.0005
20 V 0.01 + 0.003 0.015 + 0.004 0.0020 + 0.0005
200 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
1000 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
*
3
Uncertainty given as ±(% of reading + % of range).
www.tektronix.comdmm 5