8_PO-CON1692E.pdf - 第6页

Analysis of 25-OH Vitamin D2/D3 in Serum by LC-MS/MS with full-automated sample preparation First Edition: October , 2016 Conclusion • Fully Automated sample preparation procedure resulted suitable for the quantitation o…

100%1 / 6
4
Analysis of 25-OH Vitamin D2/D3 in Serum by LC-MS/MS with
full-automated sample preparation
Fig. 3 Calibration Curves (L1-L4) and MRM Chromatograms (L1)
The calibration curves showed good linearity (R^2>0.999)
over a clinical relevant range of 4.10 - 68.5 μg/L for
25-OH Vitamin D2 and 4.68 - 77.3μg/L for 25-OH
Vitamin D3 (Fig 3). The reproducibility (N=7) at three
concentrations, including LLOQ, of each compounds was
excellent (CV<6.5%). Different day reproducibility (N=7)
for 3 days at three concentrations as well (CV<7.2%)
(Table 2).
Linearity, Precision
Result and discussion
Column Temp. : 40 ºC
Injection Volume : 30 μL
[LC] NexeraX2 System
Ionization : APCI Positive
Source conditions :
Nebulizer Gas : 1.5 L/min
Interface temperature : 375 °C
Desolvation Line : 225 °C
Heat Block temperature : 250 °C
Drying Gas : Off
Scan Type : MRM
[MS] LCMS-8050
Table 1 Analytical Condition
0 10 20 30 40 50 60 Conc. Ratio
0.00
0.25
0.50
0.75
1.00
Area Ratio
2.2 2.3 2.4 2.5 2.6 2.7
0
2500
5000
7500
10000
1:395.40>269.20(+)
1:395.40>209.20(+)
25-OH Vitamin D2
r
2
=0.999
0 10 20 30 40 50 60 70 Conc. Ratio
0.00
0.25
0.50
0.75
1.00
Area Ratio
2.3 2.4 2.5 2.6 2.7
0
2500
5000
7500
10000
2:383.40>211.20(+)
2:383.40>257.40(+)
r
2
=0.999
25-OH Vitamin D3
5
Analysis of 25-OH Vitamin D2/D3 in Serum by LC-MS/MS with
full-automated sample preparation
Fig. 4 Data correlation between Automatic sample preparation (CLAM-2000) and
Manual method (n=30 human serum samples + 2 reference material, Recipe)
Comparison of 25-OH Vitamin D3 concentration between manual sample preparation (following Recipe
specications) and automated sample preparation shows good agreement as highlighted by Passing and Bablok plot
and scores (Fig. 4).
Data correlation with Manual method
Table 2 Intra-assay and Inter-assay Precision (* n=7 ; ** 3 days)
Intra-assay*
Inter-assay**
CV%
4.4%
5.2%
LOW
(5.7μg/L)
2.3%
4.7%
Medium
(17.1μg/L)
1.1%
3.7%
High
(40.0μg/L)
25-OH Vitamin D2
Intra-assay*
Inter-assay**
CV%
2.9%
5.1%
LOW
(6.3μg/L)
5.1%
5.3%
Medium
(18.9μg/L)
2.4%
5.2%
High
(43.0μg/L)
25-OH Vitamin D3
Intercept A (95%CI)
Slope (95%CI)
Corr. Coeff. (P<0,0001)
0,41
1,006
0,98
CLAM-2000
Manual Sample Preparation
Analysis of 25-OH Vitamin D2/D3 in Serum by LC-MS/MS with
full-automated sample preparation
First Edition: October, 2016
Conclusion
• Fully Automated sample preparation procedure resulted suitable for the quantitation of 25-OH Vitamin D by
elimination of all manual preparation steps.
• The automation of the method increases the analytical performance, reduces the risk for human operators and,
due to the reduced reagent consumption, reduces also the cost of the analysis.