Use as A Hand-Held Field Calibrator for Loop
Testing of Instrumentation Systems, and for
Production Facility Maintenance
Example applications
• Instrumentation system loop testing (4-20 mA)
• Validate thermostatic devices and distributors
• Use as a calibrator and low-power genera-
tor for production facilities maintenance and
electronic device evaluation
• Use as an electronic load (max. 25 mA DC)
• Simultaneously sources up to ±25 V and ±25 mA
while measuring up to ±28 V and ±28 mA (DC)
•
Eight thermoelectromotive force sourcing (TC)
modes (K, E, J, T, R, S, B, and N)
•
TEMP measurement from -25 to 80°C
•
Improved stability and lower calibration costs
•
Control from a PC via USB connection
SS7012 Features
24 V DC Source
(using the Monitor function)
SS7012
Two-Wire
Transmitter
Sensor
4-20 mA
Veri es proper sensor output (two-wire transmitter sensors). The
SS7012 applies distributor power to the sensor as a constant-
voltage (CV) output, and monitors current output from the sensor.
Loop test (two-wire transmitter sensor)
4-20 mA Current Source
SS7012
Distributor
Out
put Device
4-20 mA
Test distributor operation. Apply the default current from the
SS7012 as a substitute for sensor current output. Set SS7012 po-
larity to negative (to operate as a sink).
Loop test (distributor)
Substitute the output of sensors such as TC (thermocouple) volt-
age, current and resistance with TC, CV, CC and Rs (100 Ω) test
signals from the SS7012.
Thermometer calibration (TC, CV, and CC source)
Place Temperature
Probe near ther-
mometer terminals
Thermometer
TC (RJ) Source
SS7012
Copper Wire
or similar
Up to 20 levels can be stored for each source function, which can
be recalled manually or in sequence automatically (auto scan).
Use this function for meter checking, repetitive adjustments, and
linearity testing.
Store up to 20 steps in memory
Output
Time
Recall or Auto Scan Output
Use the optional 9184 Temperature Probe to measure tempera-
tures from -25 to 80°C. As well as measuring ambient tempera-
ture, you can calibrate thermoelectric thermometers with refer-
ence junction compensation TC (RJ).
Temperature measurement
Despite the small size and light weight, the high precision and
stability of ±0.03% for CV and CC sourcing, and ±0.05% ±0.5°C
*
for TC (0°C) sourcing, have one-year guaranteed accuracy. The
compact hand-held design is convenient for travel, eld calibra-
tion, and servicing applications.
*
For TC (0°C) sourcing of K, E, J, and T thermoelectromotive force .
Portable, precise and stable

3
Use as A Signal Source for Production Facility
Maintenance, plus Electronic Device Testing and
Evaluation.
Input
Electronic Circuit or Device
SS7012
Output
Source up to ±25.000 V and ±25.000 mA while measuring up to
±28.000 V and ±28.000 mA. Isolated source and measurement
circuits can increase operating efficiency for applications such
as simultaneous input and output testing of transducers using a
single instrument.
Simultaneous voltage and current
sourcing and display
SS7012 Sample PC Application Screen
The SS7012 PC application is available at no cost from the HIOKI
website.
USB
Cable
Use as a portable signal source for electronic device testing and
evaluation. Full external control is available via the USB inter-
face.
*
*
Only for communicating with a PC. Accessing USB memory is not supported.
Small source applications
[SOURCE FUNCTION]
Source function selection key
[TC TYPE]
Emulated thermocouple type
selection key
[MEMORY]
Memory source setting key
[SET]
Source level setting keys (for each digit)
[ +/– ]
Source polarity selection key
[OUTPUT]
Source output on/off key
[MEASURE FUNCTION]
Measurement function selection key
[0 ADJ]
Measurement zero adjustment key
[MONITOR]
Measurement/Monitor value
selection key
Operation keys
• Bar indicates selected function
• Display of thermocouple setting for TC
function
• Set level and on/off output state
Source display
• Bar indicates selected function
• Measured value or output value monitor
• Display of address and interval for
memory setting and output
Measurement display
Intuitive operation is ensured by an up/down key for each set-
ting digit, allowing quick output setup.
Easy key operations

HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
DISTRIBUTED BY
All information correct as of Jan. 5, 2009. All specications are subject to change without notice. SS7012E2-01B-03K Printed in Japan
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
1608-1610 Shanghai Times Square Office, 93 Huai Hai Zhong
Road
, Shanghai, P.R.China POSTCODE: 200021
TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
E-mail: info-sh@hioki.com.cn
Beijing Ofce :
A-2602 Freetown, 58 Dong San Huan Nan Road
Beijing, P.R.China POSTCODE: 100022
TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
E-mail: info-bj@hioki.com.cn
Guangzhou Ofce :
Room A-3206, Victory PlazaServices Center, No.103,
Tiyuxi Road, Guangzhou, P.R.China POSTCODE:510620
TEL +86-20-38392673/2676 FAX +86-20-38392679
E-mail: info-gz@hioki.com.cn
SS7012 Specifications
Sourcing &
Measurement Range
Reso-
lution
Accuracy Remarks
Sourcing
Constant Voltage (CV)
2.5V: 0 to ±2.5000 V
25V: 0 to ±25.000 V
100 µV
1 mV
±0.03% of setting ±300 µV
±0.03% of setting ±3 mV
(Sink/Source)
±25 mA
±25 mA
Constant Current (CC)
25mA: 0 to ±25.000 mA 1 µA
±0.03% of setting ±3 µA
(Sink/Source)
Approx. ±25 V
Thermoelectromotive
Force TC (0°C)
K: -174.0 to 1372.0 °C
E: -220.0 to 839.0 °C
J: -208.0 to 1108.0 °C
T: -169.0 to 400.0 °C
R: -50 to 100 °C
101 to 1768 °C
S: -50 to 100 °C
101 to 1768 °C
B: 300 to 600 °C
601 to 1820 °C
N: -113.0 to 1300.0°C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
1 °C
1 °C
1 °C
0.1 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±0.5 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±1.0 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±1.0 °C
±0.05% of setting ±2.5 °C
±0.05% of setting ±1.5 °C
±0.05% of setting ±0.5 °C
Output
Resistance
5Ω or less
IEC 584
Thermoelectromotive Force TC (RJ)
Thermocouple types, source range
and resolution are the same as for
TC (0°C)
Add the following to
TC (0°C) accuracy
±0.5 °C (@23 ±5 °C)
±1.0 °C (from 0 to 50 °C)
±2.0 °C (from -25 to 80 °C)
Using
TEMPERATURE
PROBE
9184
Memory Sourcing (RECALL, SCAN)
One type for each function: CV2.5,
CV25, CC, TC (0°C and RJ)
Max. steps: 20
Interval: 1 to 99 s
Timing
accuracy is not
specied
Standard Resistance (Rs) 100Ω
±0.2 Ω ±5 mΩ/°C
Measurement
Voltage
2.5V: 0 to ±2.8000 V
25V: 0 to ±28.000 V
100 µV
1 mV
(after zero adjustment)
±0.03% of rdg. ±300 µV
±0.03% of rdg. ±3 mV
Input Resistance
Approx. 1 MΩ
Current
25mA: 0 to ±28.000 mA 1 µA
(after zero adjustment)
±0.03% of rdg. ±3 µA
Input Resistance
Approx. 30 Ω* or less
Temperature
-25 to 80 °C 0.1 °C
±0.5 °C (@23 ±5 °C)
±1.0 °C (from 0 to 50 °C)
±2.0 °C (from -25 to 80 °C)
Using
TEMPERATURE
PROBE
9184
Monitor
Load current monitor
Terminal voltage monitor
Reference contact
temperature monitor
10 µA
10 mV
0.1 °C
±0.1% of rdg. ±20 µA
±0.1% of rdg. ±20 mV
±0.5 °C (@23 ±5 °C)
±1.0 °C (from 0 to 50 °C)
±2.0 °C (from -25 to 80 °C)
0 to ±28.00 mA
0 to ±28.00 V
-25 to 80 °C
* Including approximately 15 Ω fuse resistance
(At 23 ±5°C, accuracy guaranteed for one year)
AC ADAPTER 9445-02
(For Japan, US, and Canada. 100V to 240V AC, 9V DC / 1A output)
AC ADAPTER 9445-03
(For EU. 100V to 240V AC, 9V DC / 1A output)
TEMPERATURE PROBE 9184
(reference junction compensation)
CARRYING CASE 9380 (Holds main unit only, soft case)
CARRYING CASE 9782 (Holds options, hard case)
COMMUNICATION PACKAGE SS9000
(Includes USB cable and driver software)
Options
Recommended combination examples
SS7012 (main unit) + AC ADAPTER 9445-02 / -03 +
TEMPERATURE PROBE 9184 + CARRYING CASE 9782
DC SIGNAL SOURCE SS7012
Ordering information
Commercially available rechargeable batteries (AA Ni-MH
batteries x 4) may also be used to power the SS7012. Using
locally purchased rechargeable batteries and dedicated
battery chargers is recommended; however, HIOKI will not
be able to guarantee operating time as different rechargeable
batteries exhibit different power specifications per charge.
The SS7012 cannot be used to recharge batteries.
TEMPERATURE
PROBE 9184
TEST LEAD
9170 -
10 (Provided)
INPUT CORD
9168 (Provided)
COMMUNICATION
PACKAGE SS9000
CARRYING CASE
9380
CARRYING CASE
9782
AC ADAPTER
9445
-02
SS7012 General Specifications
Source output method: Bipolar sink/source
A/D conversion method: Delta-Sigma
Temperature coefcients (sourcing):
±(0.005% of setting + 30 µV) / °C (CV: 2.5V)
±(0.005% of setting + 0.2 mV) / °C (CV: 25V)
±(0.005% of setting + 0.2 µA) / °C (CC: 25mA)
±(0.005% of setting + 0.05 °C) / °C (TC(°C): K, E, J, T, and N)
±(0.005% of setting + 0.1 °C) / °C (TC(°C): R, S, and B)
±(0.01% of setting + 0.1 °C) / °C (TC(RJ): K, E, J, T, R, S, B, and N)
±0.005 Ω / °C (Rs)
Temperature coefcients (measuring):
±(0.005% rdg. + 10 µV) / °C (V: 2.5V)
±(0.005% rdg. + 0.1 mV) / °C (V: 25V)
±(0.005% rdg. + 0.1 µA) / °C (A: 25mA)
±0.05 °C / °C (TEMP)
±(0.01% rdg. + 1 µA) / °C (MONITOR CV: 2.5V and 25V)
±(0.01% rdg. + 1 mV) / °C (MONITOR CC: 25mA)
±0.05 °C / °C (MONITOR TC (RJ))
Zero adjustment function: within ±100 counts (V, A, and measuring)
Display: Five-digit LCD (“28000”)
Operating temperature and humidity:
0 to 40 °C, 80% RH or less (non-condensing)
I/O protection: 50 mA, fused (sourcing and measuring)
Withstand voltage: 370 Vrms AC @1 mA for 15 seconds
(source-to-measurement terminals)
Power supply: Four LR6 alkaline batteries or HR6 Ni-MH
batteries, or 9445
-02 / -03 AC ADAPTER
Continuous operating time:
(w/fully charged 2500-mAh Ni-MH batteries)
Approx. 200 min (CC: 25mA range, sourcing 20 mA into 250 Ω load)
Approx. 170 min (CV: 25V range, sourcing 25 V into 1 kΩ load)
Approx. 280 min (V: 2.5V range, input terminals shorted)
Size and weight: 104W × 180H × 58D mm, 570 g
Accessories: One INPUT CORD 9168 set, one TEST LEAD
9170
-10 set, one spare fused (F50 mA, 250 V “mizet”
fuse), four LR6 alkaline batteries