Find us at www. keysigh t.com Page 12 R NA 0.01 °C 2700 °C ± 0.25 °C S NA 0.01 °C 3500 °C ± 0.35 °C T NA 0.01 °C 550 °C ± 0.06 °C Thermistor 3 2.25 kΩ NA 0.01 °C - 80 to 150 °C ± 0.1 °C 5 kΩ NA 0.01 °C - 80 to 150 °C ± 0…

100%1 / 22
Find us at www.keysight.com Page 11
M9182A / M9183A Technical Specifications and Characteristics (continued)
Function
Lowest Range
Sensitivity
M9182A and M9183A Sensitivity (nom)
DCV
200.0000 mV
0.1 µV
ACV
200.0000 mV
0.1 µV
Resistance (M9183A)
20.00000 Ω
10 µΩ
Resistance (M9182A)
200.0000 Ω
100 µΩ
DCI (M9183A)
200.0000 nA
0.1 pA
DCI (M9182A)
2.000000 mA
10 nA
ACI
2.000000 mA
1 nA
Capacitance
1000.0 pF
0.1 pF
Temperature
Function
Type
R
0
(Ω)
Sensitivity
Range/max
temperature
1 year
23 °C ± 5 °C
M9182A and M9183A temperature accuracy (spec)
1
RTD temperature
measurement
2,3
pt385
100 Ω, 200 Ω
0.01 °C
-150 to 650 °C
± 0.06 °C
500 Ω, 1 kΩ
0.01 °C
-150 to 650 °C
± 0.03 °C
Cu
(Copper)
Less than 12 Ω 0.01 °C -100 to 200 °C ± 0.18 °C at 20 °C
± 0.05 °C otherwise
Higher than 90 Ω 0.01 °C -100 to 200 °C ± 0.10 °C at ≤ 20 °C
± 0.05 °C otherwise
Thermocouple
temperature
measurement
4,5
B
NA
0.01 °C
2200 °C
± 0.38 °C
E
NA
0.01 °C
1200 °C
± 0.035 °C
J
NA
0.01 °C
2000 °C
± 0.06 °C
K
NA
0.01 °C
3000 °C
± 0.07 °C
N
NA
0.01 °C
3000 °C
± 0.10 °C
Find us at www.keysight.com Page 12
R
NA
0.01 °C
2700 °C
± 0.25 °C
S
NA
0.01 °C
3500 °C
± 0.35 °C
T
NA
0.01 °C
550 °C
± 0.06 °C
Thermistor
3
2.25 kΩ
NA
0.01 °C
-80 to 150 °C
± 0.1 °C
5 kΩ
NA
0.01 °C
-80 to 150 °C
± 0.1 °C
10 kΩ
NA
0.01 °C
-80 to 150 °C
± 0.1 °C
1. Specifications are for one hour warm up, within one hour self-cal, aperture 0.5 sec.
2. 4-wire RTD measurement, R0 variable 1 Ω to 7 k Ω.
3. For total measurement accuracy, add temperature probe error.
4. For total measurement accuracy, add thermocouple error and cold junction compensation.
5. DMM linearization temperature range may be greater than that of the thermocouple device.
Source-Measure [(spec) unless otherwise stated)]
Parameter
Closed Loop
Open Loop
M9183A source DC voltage, measure DC voltage
DC voltage source (output)
range
+10 to -10 V +10 to -10 V
DC source sink current at 5
V output
5 mA 5 mA
DAC resolution (nom)
18 bits /12 bits
12 bits
DC voltage source accuracy
1 Year, 23°± 5° C
0.015% ± 0.0004 1.00% ± 0.04
Settling time (typ)
100 ms/1 ms
1 ms
Source resistance (nom)
200 ohms
200 ohms
M9183A source DC voltage, measure DC voltage
DC voltage source (output) range -10.000 to +10.000 V
DC current measurement range 0 to ± 20 mA
Find us at www.keysight.com Page 13
Voltage resolution (nom) 5 mV
Voltage source accuracy 1 Year,
(23 °C ± 5 °C)
1,2,3,4
2.0% ± 0.4%
Settling time (typ) 100 ms
DC current measurement accuracy 0.1% + 0.005%
Current output
Compliance voltage
Minimum level
Source accuracy 1 year,
(23 °C ± 5 °C)
1,2,3
M9183A source DC current, measure DC voltage
DC voltage measurement range
0 to ± 2.0 V
< 1.25 µA
4.2 V
10 nA
1% + 1%
< 12.5 µA
4.2 V
50 nA
1% + 1%
< 125 µA
4.2 V
100 nA
1% + 0.5%
< 1.25 mA
4.2 V
1 µA
1% + 0.5%
< 12.5 mA
1.2 V
10 µA
1% + 0.5%
1. Specifications are for one hour warm up, within one hour self-cal, slow AC filter.
2. For temperatures outside the range of 23 °C ± 5 °C, but within 0 °C to 50 °C, add 0.1 × accuracy specification per °C.
3. Repetitive reading at an aperture of 133 ms or higher.
4. If DCV source > 6 V, then add current * 4 ohms; if DCV source > 8 V, then add current * 30 ohms.