6½ Digit PXI Digital Multimeters (M9181A, M9182A, M9183A).pdf - 第20页
Find us at www. keysigh t.com Page 20 Environmental and physical characteristics Temperat ure range Operating - 10° to 55 °C Non - oper ating - 40 ° to +85 °C Relat ive h umid ity Operating t o 80% at 40 °C Full acc urac…

Find us at www.keysight.com Page 19
Frequency and period (M9182A and M9183A)
Measurement method
Direct (conventional) counting
Input impedance
1 MΩ with < 300 pF
Sensitivity (130 Hz)
.001 Hz
Totalizer (M9183A only)
Active edge polarity
Positive or negative transition
Maximum count
10,000,000,000
Allowed rate
1 to 30,000 events per second
Threshold
Set threshold DAC
Accuracy
±2 counts
Capacitance (M9182A and M9183A)
Measurement method Differential charge balance: variable currents used to stimulate dV/dt
response.
Connection type
2-wire

Find us at www.keysight.com Page 20
Environmental and physical characteristics
Temperature range Operating -10° to 55 °C
Non-operating -40 ° to +85 °C
Relative humidity Operating to 80% at 40 °C
Full accuracy to 40% RH for 41 °C to 55 °C (non-condensing)
Max operative altitude
2000m (10,000 feet)
Connectors V HI*, 2-wire Ω IN, DCV OUT
V LO*, 2-wire Ω IN, DCV OUT
I HI*, 4-wire Ω IN
I LO*, 4-wire Ω IN
Sync OUT**
External Trigger IN**
Trigger and Sync common**
Sheathed banana jack
Sheathed banana jack
Sheathed banana jack
Sheathed banana jack
DIN 7, pin 2
DIN 7, pin 7
DIN 7, pin 4
Safety & EMC Refer to Declaration of Conformity for the latest revisions of regulatory
compliance at: www.keysight.com/go/conformity
Warm-up time
1 hour
* Do not connect the V HI and V LO or I HI and I LO connectors to anything when the measurement function does not utilize
those connectors.
** M9182A & M9183A only
Physical characteristics
Dimensions
3U/1-slot PXI/Compact PCI standard
Weight
0.5 kg (1 lb.)
Power dissipation
+5 V
Total power
300 mA
1.5 W max

Find us at www.keysight.com Page 21
Configuration
Model
Description
Hardware
M9181A
PXI 6½ digit multimeter, basic features
M9182A
PXI 6½ digit multimeter, high performance
M9183A
PXI 6½ digit multimeter, enhanced performance
Software
Supported Operating Systems Microsoft Windows XP (32-bit), Microsoft Windows Vista
(32/64-bit), Microsoft Windows 7 (32/64-bit)
Standard Compliant Drivers
IVI-C, IVI-COM, LabVIEW
Supported Application Development
Environments (ADE)
Visual Studio (VB.NET, C#, C/C++), LabVIEW, MATLAB,
VEE
Keysight IO Libraries Includes: VISA Libraries, Keysight Connection Expert, IO
Monitor
Accessories
34138A
Test lead set