13紫外可见分光光度计UV-1900i Plus (1).pdf - 第4页

6 7 7 多种功能提高日常测试舒适度 辅助用户正确操作完成准确测定 * 1 休眠功能 智能助手功能 默认为2h后休眠 休眠 唤醒功能 第二天 Read y 检查校正执行 确认校正条件 执行测量 • 基于数字技术的自动化功能,包括M2M,I0T和人工智能,带来更高的测试重复性和更大的可信度。 允许系统自我监控和自我诊断,在数据获取过程中,即使没有使用者输入的情况下也能进行问题处理,实现如同在专业 人员操作下的自动运行。 日常分析和特殊应…

100%1 / 8
4 5
八种可选语言(日语、英语、中文、西班牙语、葡萄牙语、德语、法语、俄语)
该仪器配备了标准手写笔,因此可以用手指
或手写笔控制操作。
用户友好界面
一目了然的仪器状态及操作程序
友好的用户体验
Photometric
Measures the photometric value at a
single wavelength or multiple (up to
eight) wavelengths.
Spectrum
Measures a sample spectrum using
wavelength scanning.
Quantitation
Generates a calibration curve from
the measurement of standards, and
then calculates the concentrations
of unknowns.
Kinetics
Measures absorbance changes as a
function of time, and obtains the
enzymatic activity values. The
kinetics measurement method or
the rate measurement method can
be selected.
Time Course
Measures changes over time in
photometric values at a specified
wavelength.
Biomethod
Quantifies DNA or protein
concentrations.
Multiple Measurement Modes
UV-1900i Plus has six basic measurement modes, which offer the optimal solution for a variety of measurements.
* The background color of the operation screen can also be set to white.
5
Display languages are available in eight languages (Japanese, English,
Chinese, Spanish, Portuguese, German, French, Russian).
The instrument is equipped with a stylus pen.
Operation can be performed with this pen or a finger.
The UV-1900i Plus on-screen user interface includes
large, easy-to-see icons deployed on a black
background, so the instrument settings are evident at
a glance. In addition, the large, easy-to-see icons
improve intuitive understanding, which enables users
to quickly become familiar with the operations.
Furthermore, the user interface is designed to
minimize transitions between windows, so users do not
get confused during the operations. The system also
features a faster response speed by adopting a new
CPU.
Easy-to-Use Interface
Grasp the Current Status and Operating Procedures at a Glance
Refined User-Friendliness
多种测量模式
UV-1900i Plus有六种基本测定模式,可以为多种测量需求提供优化方案
*触控屏的背景也可以设置为白色
4
UV-1900i Plus的用户触控屏具有大图标,分布在黑色
背景上。因此仪器设置一目了然。此外,这些易于查看
的大图标,便于直观理解,使用户能够快速熟悉操作。
此外,用户界面经过精心设计,将窗口间的转换操作最
少化,让用户在操作过程中不会感到困惑。
光度测定
测量单个波长或多个 多至8个
波长的光度值。
光谱
使用波长扫描测量样品光谱。
定量
根据标准测量生成校准曲线,然后
计算未知样品浓度。
动力学
测量吸光度随时间的变化并获得
酶活性值。可以选择动力学测量
方法或速率测量方法。
时间进程
测量指定波长的光度值随时间的
变化。
生物法
DNA和蛋白质浓度的定量测试。
6 7
7
多种功能提高日常测试舒适度
辅助用户正确操作完成准确测定
*
1
休眠功能
智能助手功能
默认为2h后休眠
休眠
唤醒功能
第二天
Ready
检查校正执行 确认校正条件 执行测量
基于数字技术的自动化功能,包括M2M,I0T和人工智能,带来更高的测试重复性和更大的可信度。
允许系统自我监控和自我诊断,在数据获取过程中,即使没有使用者输入的情况下也能进行问题处理,实现如同在专业
人员操作下的自动运行。
日常分析和特殊应用都可以获得高质量和高重复性的数据,而无需考虑操作者的专业水平。
启动
I
数据传输模式*
4
PictBridge 功能*
5
扫码器及键盘兼容功能
*3路由器需具备无线功能
*4可向PC传输数据
*5打印机需支持PictBridge协议
*6 扫码器和键盘可使用USB接口进行连接
开机验证功能
网络连接功能
*2 该功能可选择启用或关闭。
ON
检查基线校正,自动调零校正
或样品空白校正是否执行,并
提醒用户
*1该功能可以开启/关闭
提醒用户校正是否按预设参
数正确执行
当 UV-1900i Plus 还未准备好开
始测量时,以及在开始测量和进
100% 透光率(0 Abs)校正
时,提醒用户
休眠功能指UV-1900i Plus使用后或在特定一段时间后仪器自动进入休眠状态,从而降低能耗,延长光源使用寿命
在设置的时间自动唤醒休眠状态下的UV-1900i Plus。分析者无需等待仪器预热,提高工作效率。
仪器开机后自动执行性能检查
*
2
,确保操作可靠。此功能可与
唤醒功能结合使用。
数据可以通过网络传输给PC机。
基于无线打印技术,多台UV仪器可共享一台打印机进行数据打印
*
3
。(需要安装路由器和其他网络设备)
可通过扫码器或者键盘输入样品的名称及序号。此项功能可避免多个样品分析时输入的错误,防止样品混淆,减少
人为错误的发生,节省劳动力
*
6
8 9
Note: One method of determining the
lowerlimit of quantitation is to use ten
times the standard deviation. This is an
actual measured value and is not
guaranteed.
低杂散光
High Reproducibility and Repeatability Accuracy
The photometric repeatability accuracy is 0.0002 Abs max. (0.5 Abs and 1.0 Abs). As a result, variance in the
measurement results is suppressed, enabling more accurate quantitation and the detection of low-concentration
samples.
No.
1
2
3
4
5
6
7
8
9
10
Standard
Deviation σ
Absorbance of Blank Solution (273 nm)
-0.00001
0.00001
-0.00002
0.00002
0.00001
-0.00003
0.00001
-0.00004
0.00001
0.00005
0.000025
2.5
2.0
1.5
1.0
0.5
0.0
Calibration Curve
Conc. (mg/L)
Abs.
0.0 10.0 20.0 30.0 40.0
The figure on the right is a calibration
curve for caffeine, created with
absorbance at 273 nm. The calibration
curve has an Abs = 0.0528 Conc.
The lower limit of quantitation
determined from the standard
deviation is 0.0051 mg/L.
Ultra-Fast Scan
Spectra can be acquired as fast as 29,000 nm/min. Ultra-fast scanning is effective for tracking chemical reactions in a
short time. In addition to the absorbance change at specified wavelengths, spectra can be acquired in a
short time with the UV-1900i Plus. Therefore, more detailed behavior can be investigated by observing spectra
with the UV-1900i Plus.
The figures below show the analysis of the particle agglomeration process when salts are added to silver
nanoparticles. Measurements of the 300 to 700 nm region were performed in ultra-fast scan mode. In addition
to the decrease of absorbance at 400 nm and the increase of absorbance at 480 nm, the temporal changes of
spectra can be observed.
高性能满足多样化的需求
1.0
0.8
0.6
0.4
0.2
0.0
0.97
0.92
0.87
0.77
0.72
0.67
0.62
0.23
0.21
0.19
0.17
0.15
0.13
0.11
0.09
0.07
Abs.
Absorbance at 400 nm (Abs)
Absorbance at 480 nm (Abs)
nm
300.0 400.0 500.0 600.0 700.0
480 nm
400 nm
0 100 200 300
Time (sec)
Temporal Changes of Absorbance at 400 and 480 nm
Absorption Spectra of Silver Nanoparticles
400 500 600
Black: after 0 min
Red: after 0.5 min
Blue: after 1 min
Green: after 2 min
Orange: after 5 min
Purple: after 10 min
Calibration Curve
Conc. (mg/mL)
2.00
1.80
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
Abs.
2.001.000.00
Instrument with 1.00 % Stray Light
UV-1900 series
9
Low Stray Light
高光度重现性
编号
1
2
3
4
5
6
7
8
9
10
标准偏差σ
空白溶液吸光度(273 nm)
-0.00001
0.00001
-0.00002
0.00002
0.00001
-0.00003
0.00001
-0.00004
0.00001
0.00005
0.000025
2.5
2.0
1.5
1.0
0.5
0.0
标准曲线
浓度(mg/L)
Abs.
0.0 10.0 20.0 30.0 40.0
Linearity will be lost in the high absorbance region due
to stray light. UV-1900 series is equipped with Low Stray
Light Diffraction Gratings, achieve “best-in-class” low
stray light. Stray light is at 0.5 % max. (198 nm), making
accurate measurements possible at 1 Abs or more, even
in the ultraviolet region. In addition, high-concentration
samples can be quantified accurately.
The figure on the right is a calibration curve for acetic
acid, created with absorbance at 200 nm. The correlation
coefficient is 0.9998 and correct measured values are
obtained even with high-concentration samples.
超快速扫描
High Performance to Meet Diverse Needs
1.0
0.8
0.6
0.4
0.2
0.0
0.97
0.92
0.87
0.77
0.72
0.67
0.62
0.23
0.21
0.19
0.17
0.15
0.13
0.11
0.09
0.07
Abs.
吸光度(Abs)
吸光度 (Abs)
nm
300.0 400.0 500.0 600.0 700.0
480 nm
400 nm
0 100 200 300
Time (sec)
吸光度的动力学改变(400nm及480nm处)
银纳米粒子的吸收光谱
400 500 600
黑线: 0 min
红线: 0.5 min后
蓝线: 1 min后
绿线: 2 min后
橙线: 5 min后
紫线: 10 min后
Calibration Curve
Conc. (mg/mL)
2.00
1.80
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
Abs.
2.001.000.00
Instrument with 1.00 % Stray Light
UV-1900 series
8
由于杂散光影响,高吸光度区域不符合线性关
系。UV-1900 系列配备LO-RAY-LIGH衍射光栅
实现了 “同级别” 的低杂散光。杂散光为 0.5%
(198 nm),即使在紫外区域,也能够在吸光
度为 1 及以上时进行精确测量。此外,还能对高
浓度样品进行准确的定量分析。
右图是根据乙酸在 200 nm处的吸光度绘制的校
准曲线。相关系数为 0.9998,即使是高浓度的
样品也能获得准确的测量值。
光谱的测量速度可以达到29,000nm/min。超快扫描可以在短时间内追踪化学反应过程。除了在特定波长处监测
吸光度的变化以外,UV-1900i Plus可以在短时间内获取全光谱数据。因此,可通过UV-1900i Plus测试的光谱,对
更详细的动力学行为进行研究。
下图例子显示将无机盐加入到银纳米颗粒时的粒子聚集过程分析。300-700nm的光谱采用超快模式进行扫描。除了
可以看到400nm处吸光度的降低以及480nm处吸光度的升高以外,还可测量得到整个光谱的动力学变化过程。
光度重现性为0.0002 Abs(0.5 Abs 和 1.0 Abs),如此高的光度重复性,可降低测量结果差异,同时更准确地定
并检测低浓度样品。
右图是咖啡因的校准曲线,在
273nm处产生的吸光度。校准
曲线Abs=0.0528浓度,标准偏
差为0.0051 mg/L。
备注:测定定量下限的一种方法是使用10倍标准偏
差。这是一个真实测量值,而不是保证值。