Specications All the technical specications are guaranteed when the instrument has been working for more than 30 minutes under the specied operating temperature. DC Output (0°C to 40°C) Model V oltage/Current Rating O…

Design Features
With extremely low noise, the product can satisfy your
demands for highly pure power.
Excellent line regulation rate and load regulation rate
ensure the output stability and safety.
When the timing output is enabled, the system will
congure the voltage, current, and the duration time
based on the preset timer parameters, so as to provide
varied voltage and/or current output for the load.
It supports storing and recalling state les and timer
les, and allows you to restore the instrument settings to
defaults.
You can clearly view the status of the instrument from its
intuitive user interface. The help information is displayed
in real time at the bottom of the interface, convenient for
you to operate.
You can set thresholds for OVP and OCP values. If
overvoltage or overcurrent occurs, the power supply
shuts down the output automatically, and then a prompt
message is displayed.
When multiple power supplies are in serial or parallel
connection, enabling the external trigger function can
realize the synchronous output for multiple power
supplies.
The transient response time is less than 50 μs. When the
transient change occurs to the load current, the output
voltage can be quickly restored to the set value, ensuring
the output quality.
Low ripple and noise
Excellent line regulation rate and
load regulation rate: 0.01%
Fast transient response time
Sound overvoltage/overcurrent
protection (OVP/OCP)
Powerful timing output function
Easy-to-use function of le storage
and recalling
Convenient trigger function
Clear and intuitive user interface,
easy to operate

Specications
All the technical specications are guaranteed when the instrument has been working for more than 30 minutes under the specied
operating temperature.
DC Output (0°C to 40°C)
Model Voltage/Current Rating OVP/OCP
DP711 0 V to 30 V/0 A to 5 A 0.01 V to 33 V/0.01 A to 5.5 A
DP712 0 V to 50 V/0 A to 3 A 0.01 V to 55 V/0.01 A to 3.3 A
Load Regulation, ±(% of Output + Offset)
Voltage <0.01% + 2 mV
Current <0.01% + 2 mA
Line Regulation, ±(% of Output + Offset)
Voltage <0.01% + 2 mV
Current <0.01% + 2 mA
Ripple and Noise (20 Hz to 20 MHz)
Model Normal Mode Voltage Normal Mode Current
DP711 <500 μVrms/3 mVpp
<2 mArms
DP712 <500 μVrms/4 mVpp
Annual Accuracy
[1]
(25°C ± 5°C), ±(% of Output + Offset)
Programming
Voltage 0.05% + 20 mV
Current 0.2% + 10 mA
Readback
Voltage 0.05% + 20 mV
Current 0.2% + 20 mA
Resolution
Programming
Voltage
Standard: 10 mV
High resolution option installed: 1 mV
Current
Standard: 10 mA
High resolution option installed: 1 mA
Readback
Voltage
Standard: 10 mV
High resolution option installed: 1 mV
Current
Standard: 10 mA
High resolution option installed: 1 mA
Display
Voltage
Standard: 10 mV
High resolution option installed: 1 mV
Current
Standard: 10 mA
High resolution option installed: 1 mA
Transient Response Time
Less than 50 μs for output voltage to recover to within 15 mV following a change in output current from full load to half load (or from
half load to full load).
Command Processing Time
[2]
<100 ms

OVP/OCP
Accuracy, ±(% of Output + Offset) 0.5% + 0.5 V/0.5% + 0.5 A
OVP Activation Time <10 ms (OVP>1 V)
Voltage Programming Speed
[3]
(within 1% of the total variation range)
Up
Full Load 150 ms
No Load 100 ms
Down
Full Load 30 ms
No Load 450 ms
Temperature Coefcient
[4]
, ±(% of Output + Offset)
Voltage 0.01% + 2 mV
Current 0.02% + 3 mA
Stability
[5]
, ±(% of Output + Offset)
Voltage 0.02% + 2 mV
Current 0.1% + 3 mA
Mechanical
Dimensions 140 mm (W) x 202mm (H) x 332 mm (D)
Weight Net weight: 6.9 kg
Power
AC Input Power (50 Hz to 60 Hz) 100 Vac ± 10%, 120 Vac ± 10%, 220 Vac ± 10%, and 240 Vac ± 10% (max: 253 Vac)
Maximum Input Power 400 VA
Interface
RS232 1 (Male)
Environment
Cooling Method Fan cooled
Operating Temperature 0°C to 40°C for full rated output
Maximum Output Floating Voltage to
Ground
±240 Vdc
Storage Temperature -40°C to 70°C
Humidity 5% to 80% RH
Altitude Below 2,000 m
Note
[
1
]
: The accuracy parameters are acquired through calibration under 25°C after 1-hour warm-up.
Note
[2]
: The maximum time required for the output to begin to change after receiving the APPLy and SOURce commands.
Note
[3]
: Exclude the command processing time.
Note
[4]
:
Maximum change in output/readback per °C after a 30-minute warm-up.
Note
[5]
: Following a 30-minute warm-up, change in output over 8 hours under constant load, line, and ambient temperature.