DP700_DataSheet_EN.pdf - 第6页
Three years for the mainf rame. W arranty Period Order Information Description Order No. Model Programmable Linear DC Power Supply (single channel, 30 V/5 A) DP71 1 Programmable Linear DC Power Supply (single channel, 50…

Specications
All the technical specications are guaranteed when the instrument has been working for more than 30 minutes under the specied
operating temperature.
DC Output (0°C to 40°C)
Model Voltage/Current Rating OVP/OCP
DP711 0 V to 30 V/0 A to 5 A 0.01 V to 33 V/0.01 A to 5.5 A
DP712 0 V to 50 V/0 A to 3 A 0.01 V to 55 V/0.01 A to 3.3 A
Load Regulation, ±(% of Output + Offset)
Voltage <0.01% + 2 mV
Current <0.01% + 2 mA
Line Regulation, ±(% of Output + Offset)
Voltage <0.01% + 2 mV
Current <0.01% + 2 mA
Ripple and Noise (20 Hz to 20 MHz)
Model Normal Mode Voltage Normal Mode Current
DP711 <500 μVrms/3 mVpp
<2 mArms
DP712 <500 μVrms/4 mVpp
Annual Accuracy
[1]
(25°C ± 5°C), ±(% of Output + Offset)
Programming
Voltage 0.05% + 20 mV
Current 0.2% + 10 mA
Readback
Voltage 0.05% + 20 mV
Current 0.2% + 20 mA
Resolution
Programming
Voltage
Standard: 10 mV
High resolution option installed: 1 mV
Current
Standard: 10 mA
High resolution option installed: 1 mA
Readback
Voltage
Standard: 10 mV
High resolution option installed: 1 mV
Current
Standard: 10 mA
High resolution option installed: 1 mA
Display
Voltage
Standard: 10 mV
High resolution option installed: 1 mV
Current
Standard: 10 mA
High resolution option installed: 1 mA
Transient Response Time
Less than 50 μs for output voltage to recover to within 15 mV following a change in output current from full load to half load (or from
half load to full load).
Command Processing Time
[2]
<100 ms

OVP/OCP
Accuracy, ±(% of Output + Offset) 0.5% + 0.5 V/0.5% + 0.5 A
OVP Activation Time <10 ms (OVP>1 V)
Voltage Programming Speed
[3]
(within 1% of the total variation range)
Up
Full Load 150 ms
No Load 100 ms
Down
Full Load 30 ms
No Load 450 ms
Temperature Coefcient
[4]
, ±(% of Output + Offset)
Voltage 0.01% + 2 mV
Current 0.02% + 3 mA
Stability
[5]
, ±(% of Output + Offset)
Voltage 0.02% + 2 mV
Current 0.1% + 3 mA
Mechanical
Dimensions 140 mm (W) x 202mm (H) x 332 mm (D)
Weight Net weight: 6.9 kg
Power
AC Input Power (50 Hz to 60 Hz) 100 Vac ± 10%, 120 Vac ± 10%, 220 Vac ± 10%, and 240 Vac ± 10% (max: 253 Vac)
Maximum Input Power 400 VA
Interface
RS232 1 (Male)
Environment
Cooling Method Fan cooled
Operating Temperature 0°C to 40°C for full rated output
Maximum Output Floating Voltage to
Ground
±240 Vdc
Storage Temperature -40°C to 70°C
Humidity 5% to 80% RH
Altitude Below 2,000 m
Note
[
1
]
: The accuracy parameters are acquired through calibration under 25°C after 1-hour warm-up.
Note
[2]
: The maximum time required for the output to begin to change after receiving the APPLy and SOURce commands.
Note
[3]
: Exclude the command processing time.
Note
[4]
:
Maximum change in output/readback per °C after a 30-minute warm-up.
Note
[5]
: Following a 30-minute warm-up, change in output over 8 hours under constant load, line, and ambient temperature.

Three years for the mainframe.
Warranty Period
Order Information
Description Order No.
Model
Programmable Linear DC Power Supply (single channel, 30 V/5 A) DP711
Programmable Linear DC Power Supply (single channel, 50 V/3 A) DP712
Standard Accessories
Power Cord -
Either one of the following specied fuses:
Fuse 50T-050H 250V 5A (AC Selector: 100 Vac or 120 Vac)
Fuse 50T-025H 250V 2.5A (AC Selector: 220 Vac or 240 Vac)
-
Quick Guide (hard copy) -
Optional Accessories
High Resolution HIRES-DP700
Trigger (external synchronous trigger input and output) TRIGGER-DP700
Timer TIMER-DP700
9-Pin RS232 Cable (female-to-female, straight) CB-DB9-DB9-F-F-150
DP700 Series Rack Mount Kit (for a single instrument) RM-1-DP700
DP700 Series Rack Mount Kit (for two instruments) RM-2-DP700
DP700 Series Rack Mount Kit (for three instruments) RM-3-DP700
HEADQUARTER
RIGOL
TECHNOLOGIES, INC.
No.156,Cai He Village,
Sha He Town,
Chang Ping District, Beijing,
102206 P.R.China
Tel:+86-10-80706688
Fax:+86-10-80705070
Electronic Measurement
Instrument service and support
email:EMD_support@rigol.com
EUROPE
RIGOL
TECHNOLOGIES GmbH
Lindbergh str. 4
82178 Puchheim
Germany
Tel: 0049- 89/89418950
Email: info-europe@rigoltech.com
NORTH AMERICA
RIGOL
TECHNOLOGIES,
USA INC.
10200 SW Allen Blvd, Suite C
Beaverton, OR 97005, USA
Toll free: 877-4-RIGOL-1
Ofce: (440) 232-4488
Fax: (216)-754-8107
Email: info@rigol.com
JAPAN
RIGOL
TECHNOLOGIES JAPAN G.K.
Tonematsu Bldg. 5F, 2-33-8 Nihonbashi-
Ningyocho, Chuo-ku,
Tokyo 103-0013
Japan
Tel: +81-3-6264-9251
Fax: +81-3-6264-9252
Email: info-japan@rigol.com
RIGOL
®
is the registered trademark of
RIGOL
Technologies, Inc. Product information in
this document subject to update without notice. For the latest information about
RIGOL
's
products, applications and services, please contact local
RIGOL
ofce or access
RIGOL
ofcial website:
www.rigol.com
April 2016