9 www .nordsonefd.com info@nordsonefd.com USA & Canada 800-556-3484 Europe +44 (0) 1582 666334 Asia +86 (21) 3866 9006 Sales and service of EFD dispensing systems are available worldwide. Unpacking Unit Unpacking 1. …

100%1 / 16
8
www.nordsonefd.com info@nordsonefd.com USA & Canada 800-556-3484 Europe +44 (0) 1582 666334 Asia +86 (21) 3866 9006
Sales and service of EFD dispensing systems are available worldwide.
Classifications and Specifications
The ProcessMate PM500 Centrifuge is classified under “Class I” equipment.
CE Marking
The manufacturer hereby declares the conformity with the actual required safety standards in according
with 93/42 EEC and 93/68/EEC
Standards
The equipment complies with EC Electromagnetic Compatibility requirement of EN60601-1: 1990
+A13: 1996. Part 1: General requirement for SAFETY. EMI filter is designed in conformity with 89/336
EEC-EMC Directive. The equipment complies EN standard.
• ISO9001, ISO13485
• GMP (Good Manufacturing Practice)
• FDA (USA)
Specifications
Part Number 7015542
RoHS compliant Yes
Overall dimension 320 x 400 x 310mm approx.
New weight 16.0 kg (35.2 lbs) approx.
Speed range 0 – 5000 rpm adjustable, LED display
Timer 2-digits, 99 min. LED display
Chamber Stainless steel
Motor 80 W
Speed control Microprocessor
Imbalance detection Yes, alarm & brake
Errors detection Yes (4 errors detection)
Brake Automatic braking system
Safety Interlocking lid Yes
Lid open override device Yes, manual
Safety shutout device Yes
Over heat alarm Yes
Emergency STOP Yes
Power source Multi-voltage 100V ~ 240V, 50/60Hz
Quality Standard ISO9001, ISO13485, CE, GMP
9
www.nordsonefd.com info@nordsonefd.com USA & Canada 800-556-3484 Europe +44 (0) 1582 666334 Asia +86 (21) 3866 9006
Sales and service of EFD dispensing systems are available worldwide.
Unpacking Unit
Unpacking
1. Make sure the packaging is laid out in the correct orientation before opening the box.
2. Remove all packaging material around the PM5000 Centrifuge unit before attempting to remove unit
from box.
3. Carefully remove The PM5000 Centrifuge unit from the box.
4. Check unit for damage before discarding shipping packaging and box.
5. Should any damage be detected, please call your local representative of EFD Technical Services for
assistance.
Caution
1. Be careful not to drop the PM5000 Centrifuge, or place / rest the unit upside down. Unit is
heavy and you may require assistance in lifting the PM5000 Centrifuge unit.
2. Do not place the PM5000 Centrifuge in the high draft are, direct sunlight or near equipment
emitting heat and / or electromagnetic conduction.
3. Conditions for transportation & storage:
Temperature 10ºC – 40ºC
Humidity 40% 90%
4. Check the specification label located on back of this equipment, the power source must
conform to your local codes.
10
www.nordsonefd.com info@nordsonefd.com USA & Canada 800-556-3484 Europe +44 (0) 1582 666334 Asia +86 (21) 3866 9006
Sales and service of EFD dispensing systems are available worldwide.
Set-up
Placement Surface
The PM5000 weighs approximately 35.2 Lbs. / 16 Kg, and will be spinning at high speeds. Select a
level, sturdy and stable surface to place the PM5000 Centrifuge unit before use.
Power Source
The ProcessMate PM5000 is a multi voltage unit with a built in power converter. The unit is capable and
will operate in the ranges of 100 – 240 VAC 50/60 HZ. Locate the power cord shipped with the unit,
plug one end into the unit and the other to a suitable power source. Make sure the ground pin on the
power cord is connected to ground on the power source.