index-1.php.pdf - 第12页
12 D39051 | Date 12/2019 | v2.9 AtmoCONTROL FDA Edition – IQ-OQ-Documentation User Rights Login as “T estUser1”. Check if it is not possible to change the system settings ( Options → Email options or Edit backup folder )…

D39051 | Date 12/2019 | v2.9 11
AtmoCONTROL FDA Edition – IQ-OQ-Documentation
Create a new user account with the same name as before. Does the
following message appear?
yes no
Can the admin reset the password of “TestUser1” ?
yes no
Close the User Management window.
Click the Switch user button in the Security menu. Can you login as
“TestUser1”?
TestUser1
********
yes no
Is “TestUser1” displayed in the title bar?
yes no

12 D39051 | Date 12/2019 | v2.9
AtmoCONTROL FDA Edition – IQ-OQ-Documentation
User Rights
Login as “TestUser1”.
Check if it is not possible to change the system settings
(Options → Email options or Edit backup folder)
Is it impossible?
yes no
Check if it is not possible to open the User management
(Security → Edit user)
Is it impossible?
yes no
Audit Trail
Login as Admin.
To confirm the functionality of the Audit Trail function, select the
Security tab on top of the tool bar and click on Audit Trail.
Is the Audit Trail window below displayed?
yes no
Does the Audit Trail window appear without any error messages?
yes no
Activate Select time span. Is it possible to select the current day as
time span?
yes no

D39051 | Date 12/2019 | v2.9 13
AtmoCONTROL FDA Edition – IQ-OQ-Documentation
Click the Show audit trail button. Does it open up the following
browser window?
yes no
3.3 Connect database
To confirm the functionality of the Connect database function, click
Device → Connect offline from Database. Does the following win-
dow appear?
Connect demo device Y6160011.atdbx .
yes no