CP40 service manual.pdf - 第86页
1. Head Part (CP-40) WA COMP ASS'Y (Adjustment Procedures) # Connect J1 and Mother Board. 1) Set Control SW(SW1) to UP. 2) Press SW2 to RESET. 3) Open C:\SACOM File. 4) Perform Light ON OFF TEST for Aligner at F unc…

1. Head Part
(CP-40)
200 SET
RESET
SET
Ripple Control
Control SW
Connecting Cable
0 SET
BOARD
SW2 VR3 SW1 VR2 VR1
No.1
No.2 No.3
(Fig. 1)
J1
1-3-3. Align Set_Up
Version Date WA QA CA Note
00 Nov04 O

1. Head Part
(CP-40)
WA
COMP ASS'Y
(Adjustment Procedures)
# Connect J1 and Mother Board.
1) Set Control SW(SW1) to UP.
2) Press SW2 to RESET.
3) Open C:\SACOM File.
4) Perform Light ON OFF TEST for Aligner at Function.
5) Align COMP ASS'Y with Aligner Bottom.
# IMAGE DROW => Profile Open
6) Adjust VR2 to Set Waveform to 0.
7) Adjust VR3 to Set Ripple Waveform to Some.
8) Adjust VR1 to Set Waveform to 200.
Version Date WA QA CA Note
00 Nov04 O O O

1. Head Part
(CP-40)
(Fig. 3)
(Adjustment Procedures)
9)
Using a Fixture(with Bolt Mark Inside), Align Aligner with WA Light.
10) At IMAGE => EDGE, Raise the Fixture Up to Indicate Figures on Both Sides(Refer
to Fig. 2).
11) While Rotating R-Axis, Make Five Digits on Both Sides Indicated Almost Similarly.
12) Download File to Save Average Distance(Rounding Off the Numbers to Three
Digits) as shown in Fig. 2 .
Example)
Average Distance = 115.31 => 5.3
13) Press SW2 Again and Set SW1 to Down.
ALIGNER
Version Date WA QA CA Note
00 Nov04 O O O