SW1002_user_manual_chi_20191017H.pdf - 第108页
104 (7) E X T. I / O EXT .I/O :IO:FILTer:STATe <1/0/ON/OFF> :IO:FILTer:STATe? <1/0> 1/ON 0/OFF EXT .I/O SCAN …

103
(6)
RS-232C
:SYSTem:COMMunicate:FORWard:RS232C:SPEED
<9600/19200/38400>
:SYSTem:COMMunicate:FORWard:RS232C:SPEED?
<9600/19200/38400>
RS-232C
RS-232C = 8 = 1 =
=
:SYSTem:COMMunicate:FORWard:TIMeout <1
-
100/MIN/MAX/
DEF>
:SYSTem:COMMunicate:FORWard:TIMeout?
<1 100 (NR1)> [s] MIN 1, MAX 100, DEF 10
:SYST:COMM:FORW:TIM 5
5
:SYST:COMM:FORW:TIM?
30
30
A
:A <"
">
< >
RS-232C
PC
“
:A
:A ":FUNC RV"
A
:FUNC RV
:A ":READ?"
289.68E-3, 1.3921E+0
A
:READ?
PC
:A:INIT
:A:READ?
8

104
(7) EXT.I/O
EXT.I/O
:IO:FILTer:STATe <1/0/ON/OFF>
:IO:FILTer:STATe?
<1/0>
1/ON
0/OFF
EXT.I/O SCAN SCAN_RESET
/
:IO:FILT:STAT 1
:IO:FILT:STAT?
0
EXT.I/O
:IO:FILTer:TIME <0.05
0.50/MIN/MAX/DEF>
:IO:FILTer:TIME?
<0.05 0.50 (NR2)> [s] MIN 0.05, MAX 0.50, DEF 0.05
:IO:FILT:TIME 0.1
0.1 s
:IO:FILT:TIME?
0.05
0.05 s
CLOSE
:IO:PULSe:TIME <0.001
0.100/MIN/MAX/DEF>
:IO:PULSe:TIME?
<0.001 0.100 (NR2)> [s]
(MIN
0.001, MAX 0.100, DEF 0.005)
EXT.I/OCLOSE
:IO:PULS:TIME 0.001
CLOSE 0.001 s
:IO:PULS:TIME?
0.005
CLOSE 0.005 s

105
(8) RS-232C
RS-232C
:SYSTem:COMMunicate:RS232C:SPEED <9600/19200/38400>
:SYSTem:COMMunicate:RS232C:SPEED?
<9600/19200/38400>
RS-232C
(DFLT/USER) DFLT9600 bps
RS-232C = 8 = 1 =
=
(9) LAN
:SYSTem:COMMunicate:LAN:CONTrol <1
9999>
:SYSTem:COMMunicate:LAN:CONTrol?
<1 9999 (NR1)>
LAN TCP/IP
:SYSTem:COMMunicate:LAN:UPDate
(DFLT/USER) DFLT23
SYST:COMM:LAN:CONT 23
SYST:COMM:LAN:UPD
SYST:COMM:LAN:CONT?
23
:SYSTem:COMMunicate:LAN:GATeway <
1>,<
2>,<
3>,<
4>
:SYSTem:COMMunicate:LAN:GATeway?
< 1>,< 2>,< 3>,< 4>
<
1> = 0 255 (NR1)
<
2> = 0 255 (NR1)
<
3> = 0 255 (NR1)
<
4> = 0 255 (NR1)
1 4
:SYSTem:COMMunicate:LAN:UPDate
(DFLT/USER) DFLT0.0.0.0
SYST:COMM:LAN:GAT 192,168,0,100
SYST:COMM:LAN:UPD
SYST:COMM:LAN:GAT?
192,168,0,100
8