10 2. Press the “BAR” icon to chang e pressure unit to BAR 3. Touch the X icon to save and exit 5.2.3 Manual/Purge Dispense Cy cle Setting: 1. Touch the Purge icon to select the purge cycle. The Purge icon will turn to g…

100%1 / 16
24
13. Go to any computer and open any web browser
14. Type in:
http://localhost:8090/ then press the enter button
You are now in the webserver site. The below screen will be displayed:
15. Click the “Login” button
16. Enter “admin” in the Login box
17. Enter “1001” in the Password box
9
5.2 DISPENSING
5.2.1 Pressure Adjustment
Note: Pressure in Port A and Port B delivers the same output pressure.
This pressure is to activate the valve. Make sure the supplied pressure is set
at least 70 psi (4.8 bar)
1. Touch the “Pressure A.B” icon to enter the pressure setup screen
2. Touch the Up and Down arrows to set valve activation pressure (70 psi
minimum)
3. Touch the Check Mark icon to save and exit
5.2.2 To Change Pressure Unit Display
Note: The default pressure unit is PSI. To change pressure unit to BAR,
follow the instructions below.
1. Touch the Setup icon to enter the setup screen
10
2. Press the “BAR” icon to change pressure unit to BAR
3. Touch the X icon to save and exit
5.2.3 Manual/Purge Dispense Cycle Setting:
1. Touch the Purge icon to select the purge cycle.
The Purge icon will turn to green color.
2. Press and hold down the foot switch to activate the purge dispense cycle.
Alternately, touch and hold the Run icon on the display to activate the
purge dispense cycle.
5.2.4 Automatic Dispense Cycle Setting
1. Touch the “Timed” mode icon to set the dispense cycle in pre-set dispense
time. The icon should turn to green color.
2. Touch the Dispense icon to enter the setup screen
3. Touch the Up and Down arrows to set the desired dispense time.
Note: The minimum activation time is 0.008 second
23
9. Type: “set ZONEINFO=C:\iot\timezone.zip” then press the “Enter” button
10. Type: “ipconfig” then press the “Enter” button
11. Record the Server IPv4 address show in the “IPv4 Address” line on the above
screen; For this example, the Server IP address is: 172.16.40.91
This address will be entered in the controller screen later.
12. Type: “.\portal” then press the Enter button
22
7. Type: “cd :\iot” then press the “Enter” button
8. Type: “set OKI_IOT_PREFIX=MyPrefix” then press the “Enter” button
11
4. Touch the Check Mark icon to save and exit
5. Press the foot switch to activate the “Timed” dispense cycle.
Alternately, touch the run icon on the display to activate the “Timed”
dispense cycle.
Note: The unit has an “Interrupt” mode feature. In this mode, the
“Timed” dispense cycle can be disrupted if the foot switch is released and
resumed when the Foot Switch is depressed again.
6. Touch the “Interrupt” icon to activate “Interrupt” mode. The icon should
turn to green color.
5.2.5 Teach Mode Setting
In the teach mode, the dispense time will be accumulated as long as the
foot switch is depressed. This is helpful in determining the required
dispense time when the dispense output is known.
1. Touch the “Teach” icon to enter the teach mode
2. Touch the “Time Reset” icon to set timer to zero
3. Press and hold down the foot switch, the dispense time will be accumulated
4. Release the foot switch when the desired amount of fluid has dispensed
5. Touch the “Timed” icon to transfer the accumulated dispense time to
“Timed” mode
6. The unit is now set to repeat this “Timed” cycle
5.2.6 To run in Continous Mode
The controller can be setup to repeat the run continuously.
1. Touch the Setup icon to enter the setup screen