CAUTION: A 5-micron filter (TSD800-6) must be installed with the unit to ensure proper
air filtration.
Figure 2.0
5.1 Connecting the Unit: (Refer to Figure 1.0 and 2.0)
1. Connect the power cord, foot switch to the back of the unit.
2. Connect Valve air hose to either Port A or Port B
3. Press the Power switch to turn on the unit.
Items Description
1 Air Filter
2 Valve (not included)
3 Pressure Pot (not included)
4 Display
5 Power Adapter
6 Foot Switch
27
Proceed to make changes
Click “OK” button to save
The new parameters will be displayed on the controller
7. SOFTWARE UPGRADE
Note: For future software upgrade, follow the instructions below.
1. Download the latest software version from Techcon website and copy it to
a blank USB thumb drive. Note: Do not put the software file inside a
folder
2. Make sure that the unit is turned off
3. Insert the USB drive to the USB port located in the back of the unit
4. Turn on the unit
5. Wait while the system is loading the new software. This process can take
a few minutes
“Edit” icon
26
6.2 Making Parameters Adjustment from the websever
Go back to the webserver site and click on the “Controllers” menu.
Click on the active controller link
The controller screen will be displayed in the webserver as shown below.
The controller parameters such as pressure and dispense time can be adjusted
from the webserver.
Click on the “Edit” icon
The parameter screen will appear as shown below:
7
5.1.1 Login
1. Touch the Login icon to enter the login screen
2. Enter “0000” in Password window
3. Touch the Check Mark icon to save and exit
5.1.2 Pressure Calibration
Note: Pressure calibration must be performed when the unit is activated for
the first time.
8
4. Touch the Setup icon to enter the setup screen
5. Touch the Calibration icon to enter the calibration screen
Note: Make sure the air pressure line is not connected to the unit at this time.
6. Touch the “0” icon to set the pressure to 0
7. Now connect the supplied air pressure line to the back of the unit and turn
up input air pressure to 100 psi
8. Touch the “100” icon to set the pressure to 100
Note: The digital values shown at “0” and “100” icons are for reference only.
The actual calibrated values will be different.
9. Touch the Reboot icon to save the settings and reboot the system
10. Wait until the system completes the rebooting sequence and the home
screen is displayed
The unit is now calibrated and ready to operate
25
18. Go to the setup screen of the controller and touch the “Remote Server”
icon
19. Enter the Server IPv4 address recorded in step #11 into Remote IP address box
20. Touch the Check Mark icon to save
21. The controller is now connected to the webserver site.