00196624-04_Service Manual WPC5_6_EN_01-2019.pdf - 第104页

Service Ma nual W PC5 / WPC6 ➢ Fit the cover back o n the loading u nit and screw tight . See a l so... 3.5 .2 Re place the Load Axis Dr ive Mot or [ ➙ 3-3 4 ] ➢ Unscrew the se nsor (1) fro m its mount. ➢ Screw the s ens…

100%1 / 143
Service Manual WPC5 / WPC6
3.7.14 Sensor 15 NSM-Modul (WPC6) Safety flap open
Spare Part
Sensor safety flap open [03056958-xx]
Removal / Installation
Remove the cover on the load unit (see "3.5.1.1 Remove the Cover on the Load Unit" [ 3-
32]).
Loosen the screws marked at 1 to 4
and lift off the cover.
Disconnect connector X3 for the
safety flap when open from the
circuit board.
Service Manual WPC5 / WPC6
Fit the cover back on the loading unit and screw tight.
See also...
3.5.2
Replace the Load Axis Drive Motor [ 3-34]
Unscrew the sensor (1) from its
mount.
Screw the sensor into its mount.
Use a feeler gauge to set the
distance between the safety flap
and the sensor to 0.2 mm.
Reconnect connector X3 to the
circuit board.
Service Manual WPC5 / WPC6
3.7.15 Sensor 17 NSM (WPC6) Safety Sensor loading flap
Spare Part
Safety sensor loading flap [03056989-xx]
Removal / Installation
Remove the cover on the load unit (see "3.5.1.1 Remove the Cover on the Load Unit" [ 3-
32]).
Open the loading flap.
Disconnect the cable x2j from the sensor and unthread it.
Remove the two fastening screws and then take off the sensor.
Insert the new sensor and fix into place with the two screws.
Thread the sensor cable into place again and reconnect it.
Fit the cover back on the loading unit.
Legend
1. Actuator BPS 260-2
2. Button with signal light
3. Safety sensor loading flap