141-400-57-probestationaccessoriescatalog.pdf_-_141-400-probestationaccessoriescatalog_1.pdf.pdf - 第44页
PN 141-400 www.formfactor.com Lab Accessories and Miscellaneous • 28 Revision 57, February, 2023 Probe Station A ccessories 146 11 7 — T abl e f or SE7 5 0 Feat ur es • T a ble f or SE750 Sh ield Enclosure • Heav y duty …

PN 141-400 www.formfactor.com Lab Accessories and Miscellaneous • 27
Revision 57, February, 2023 Probe Station Accessories
– Optimum load on the tabletop is 90-363 kg (200-800 lb)(since
the pressure is directly related to the mass sitting on the
tabletop, the operating pressure determines this range)
– Vibration isolation: typical performance better than 0db at 6 Hz,
with –6 dB per octave roll-off to 48 Hz and >–18 dB attenuation
above 48 Hz (with a Summit 12000 seri es probe station or
equivalent load)
Compatibility
• Summit, M150, BTS
177-396 — Vibration Isolation Table, DELUX Package for
200 mm Stations
Features
• Heavy duty frame and steel tabletop
• Integrated rolling casters
• Seismic restraint kit
• Storage s helf below tabletop
• Ergonomic front support bar
• Optional side instrument she lves and
accessories available
• Computer accessory mounting kit for
keyboard, mouse, joystick
• Multi-directional LCD monitor supp ort arm (VESA).
• Enclosed system (side walls and front access door).
Specifications
• Dimensions:
– Steel tabletop (W x D): 1016 x 1016 mm (40 x 40 in)
– Table operating height: 813 mm (32 in)
• Weights:
– Steel tabletop:104 kg (230 lb)
– Complete system (including tabletop): ~218 kg (~480 lb)
• Operation:
– Optimum air p ressure: 25 - 60 psi
– Input air pressu re: min 20 psi (approx.) - max 80 psi
– Max weight capaci ty 703 kg (1,150 lb) (@ 80 psi)
– Optimum load on the tabletop is 90-363 kg (200-800 lb) (since
the pressure is directly related to the mass sitting on the
tabletop, the operating pressure determines this range)
– Vibration isolation: typical performance better than 0 db at 6 Hz,
with –6 dB per octave roll-off to 48 Hz and >–18 dB attenuation
above 48 Hz (with a Summit 12000 seri es probe station or
equivalent load)
Compatibility
• Summit
143-567 — Side Instrument Shelf for Table
Features
• Universal design can mount on left or right
side of table
• Heavy duty design can hold multiple
instruments
• Easy mounting enables quick reconfiguration
Specifications
• Dimensions:
– Side shelf (W x D): 533 x 1016 mm (21 x 40 in), positioned with
a 25.4 mm (1 in) space between the side shelf and main steel
tabletop
– Table width (with 2 si de shelves installed, left and right): 2184.4
mm (86 in)
169-130 — Vibration Isolation Platform VIP601
Features
• Compact, vibration dampening
platform
• Suitable for MPS150 and EPS150
probe stations
Specifications
• Dimensions (WxDxH): 598 x 530 x 87 mm (23.5 x 20.9 x 3.4 in)
• Max. load at 4 bar : 530N per damper
• Max load capacity at nominal pressure (4 bar): 200 kg
• Internal frequency: 2.5 Hz (vert.), 2.5 Hz (hor.), depending o n load
Compatibility
• MPS150, EPS150 and also with SE750
171-512 — Optional Cover and Earthquake Kit for
VIP601
Features
• Cover for VIP601
• Earthquake protection
Specifications
• Fixed link between pr obe station
and p latform p rovides seismic stability
• Enables probe stat ion movement together with the platform in case
of relocation or a seismic event
Compatibility
• VIP601
• Not configurable with EPS-ACC-SE750-COAX and EPS-ACC-
SE750-TRX (earthquake protection already included)

PN 141-400 www.formfactor.com Lab Accessories and Miscellaneous • 28
Revision 57, February, 2023 Probe Station Accessories
146117 — Table for SE750
Features
• Table f or SE750 Shield Enclosure
• Heavy duty frame
• Integrated rolling casters
Specifications
• Dimensions (WxDxH): 720 x 668 x
653 mm (28.3 x 26.3 x 25 .7 in)
Compatibility
• Shield En closure SE750 is
required
• Configuration with VIP601 requires SE750
• Configuration with station requires SE750

PN 141-400 www.formfactor.com Lab Accessories and Miscellaneous • 29
Revision 57, February, 2023 Probe Station Accessories
Shield Enclosures/Dark Boxes
111650 — Shield Enclosure, SE1200 EMC
Features
• For standard probe systems PM300 and
PA300 (others on request).
• Light-tight, electrically grounded and
optimized design for EMC shielding
• Vizor opening
• Requires additional feed-throughs and
table adaptation
• Requires EMO option in combination with (semiautomatic) PA300.
• Inside light and interlock switch optional
• Clean room compliant field upgrades a re not possible
• Four people a re required to complete field upgrade installation
Specifications
Required floor s pace:
• Width = 1.5 m (59 in)
• Depth = 1.3 m (51.2 in)
• Height (with table): 2.1 m (82.7 in)
Compatibility
• PA300, PM300
138121 — Shield Enclosure, SE1000 EMC with Interlock
and Illumination
Features
• For standard probe systems PM5
(with laser cutter), PM8 and PA200
(others on request).
• Light-tight, electrically grounded and
optimized design for EMC shielding
• With lamps inside and interlock switch
at front doo r
• Requires additional feed-throughs
and table adaptation
• Requires EMO option in combination with (semiautomatic) PA200
• Clean room compliant field upgrades a re not possible
• Four people a re required to complete field upgrade installation
Specifications
Required floor s pace:
• Width = 1.3 m (51.2 i n)
• Depth (with front lid opened) = 2 m (78.7 in)
• Height (with table and front lid opened): 2 .5 m (98.4 in)
• Requires:
– Adaptation to VIT70x (PN 138 390)
– VIT700 (PN 138815) or VIT701 (PN 135100) for BlueRay
– Adaptation SE1000EMC to VIT801 (PN 119585)
– Vibration isolation table VIT801 (PN 115405)
Options:
• 100689 EMC light-tight feed-through
• 112392 EMC test equipment interface, standard
• 112391 EMC test equipment interface, Kelvin
Compatibility
• PA200, PA200 BlueRay, PM8, MPS150, EPS150
100689 — EMC Light-Tight Feedthrough
Features
• For SE750EMC, SE1000EMC and SE1200EMC.
• Electromagnetically shielded
• For probe station, thermal chuck and video wiring
Compatibility
• Shield enclosures: SE750, SE1000, SE1200
112391 — EMC Test Equipment
Interface, Kelvin
Features
• For test equipment adaptation (e.g.,
HP4142/4156B)
• 4 triax-triax-dual (Kelvin) and 8 coax-coax
feedthroughs
Compatibility
• Shield enclosures: SE750, SE1000, SE1200
112392 — EMC Test Equipment
Interface, Standard
Features
• For adapting test equipment
• One plate with 8 triax-triax and 8
coax-coax feedthroughs
Compatibility
• Shield enclosures: SE750, SE1000,
SE1200