E4416A_E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors.pdf - 第20页
Find us at www.keys ight.com Page 20 Table 14. Rise and fall times v ersus sensor b andwidth. 1 Overshoot in respon se to power steps wi th fast rise times, i.e. less than the se nsor rise time, is < 10 % . When a pow…

Find us at www.keysight.com Page 19
Settling Times
Average-only mode
In normal and x 2 speed, manual filter, 10 dB decreasing power step refer to Table 12.
Table 12. Settling time (average only mode).
Number of average 1 2 4 8 16 32 64 128 256 512 1024
Settling time(s) normal 0.08 0.13 0.24 0.45 1.1 1.9 3.5 6.7 14 27 57
Settling time(s) x 2 0.07 0.09 0.15 0.24 0.45 1.1 1.9 3.5 6.7 14 27
In fast speed, within the range –50 to +20 dBm, for a 10 dB decreasing power step, the settling time is
10 ms (for the E4416A) and 20 ms (for the E4417A).
When a power step crosses the power sensor’s auto-range switch point, add 25 ms.
Normal mode
In normal, free run acquisition mode, within the range –20 to +20 dBm, for a 10 dB decreasing power
step, the settling time is dominated by the measurement update rate and is listed in Table 13 for various
filter settings.
Table 13. Settling time (normal mode).
Number of averages 1 2 4 8 16 32 64 128 256 512 1024
Settling time free run
acquisition, normal speed (s)
0.1 0.15 0.25 0.45 0.9 1.7 3.3 6.5 13.0 25.8 51.5
Settling time free run
acquisition, X2 speed (s)
0.08 0.1 0.15 0.25 0.45 0.9 1.7 3.3 6.5 13.0 25.8
In normal mode, measuring in continuous or single acquisition mode, the performance of rise times, fall
times and 99% settled results are shown in Table 14. Rise time and fall time specifications are for a
0.0 dBm pulse, with the rise time and fall time measured between 10% to 90% points and upper range
selected.

Find us at www.keysight.com Page 20
Table 14. Rise and fall times versus sensor bandwidth.
1
Overshoot in response to power steps with fast rise times, i.e. less than the sensor rise time, is < 10%.
When a power step crosses the power sensor’s auto-range switch point, add 10 µs.
1 Rise and fall time specifications are only valid when used with the E9288A sensor cable (1.5 meters).
Sensor mode, parameter Video bandwidth setting
Low Medium High Off
E9321A
E9325A
rise time (< µs)
2.6
1.5
0.9
0.3
fall time (< µs)
2.7
1.5
0.9
0.5
Settling time (rising) (< µs)
5.1
5.1
4.5
0.6
Settling time (falling) (< µs)
5.1
5.1
4.5
0.9
E9322A
E9326A
rise time (< µs)
1.5
0.9
0.4
0.2
fall time (< µs)
1.5
0.9
0.4
0.3
Settling time (rising) (< µs)
5.3
4.5
3.5
0.5
Settling time (falling) (< µs)
5.3
4.5
3.5
0.9
E9323A
E9327A
rise time (< µs)
0.9
0.4
0.2
0.2
fall time (< µs)
0.9
0.4
0.2
0.2
Settling time (rising) (< µs)
4.5
3.5
1.5
0.4
Settling time (falling) (< µs)
4.5
3.5
2
0.4

Find us at www.keysight.com Page 21
This information is subject to change without notice. © Keysight Technologies, 2014 - 2022, Published in USA, September 15, 2022, 5980-1469E
Learn more at: www.keysight.com
For more information on Keysight Technologies’ products, applications or services,
please contact your local Keysight office. The complete list is available at:
www.keysight.com/find/contactus
Physical Specifications
Dimensions
150 mm L x 38 mm W x 30 mm H (5.9 in x 1.5 in x 1.2 in)
Weight
Net: 0.2 kg (0.45 lbs)
Shipping
0.55 kg (1.2 lbs)
Mechanical
characteristic
Mechanical characteristics such as center conductor protrusion and pin depth are not
performance specifications. They are, however, important supplemental characteristics
related to electrical performance. At no time should the pin depth of the connector be
protruding
Ordering Information
E9321A
50 MHz to 6 GHz; 300 kHz BW
E9322A
50 MHz to 6 GHz; 1.5 MHz BW
E9323A
50 MHz to 6 GHz; 5 MHz BW
E9325A
50 MHz to 18 GHz; 300 kHz BW
E9326A
50 MHz to 18 GHz; 1.5 MHz BW
E9327A
50 MHz to 18 GHz; 5 MHz BW
Power Sensor Options
E932xA-A6J Supplies ANSI/NCSL Z540-1-1994 test data including measurement
uncertainties
E932xA-1A7