E4416A_E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors.pdf - 第5页
Find us at www.keys ight.com Page 5 Table 2. Measur ement speed for different sensor types. Channel functi ons A, B, A/B, B/A, A - B, B - A and Relative Storage regist ers 10 instrument stat es can be saved via the S ave…

Find us at www.keysight.com Page 4
Accuracy
Instrumentation
Please add the corresponding power sensor linearity percentage; see Tables 6a and 6b for the E9320
sensors.
Average only mode
Absolute
Logarithmic
± 0.02 dB
Linear
± 0.5%
Relative
Logarithmic
± 0.04 dB
Linear
± 1.0%
Normal mode Calibration temperature
1
± 5 °C Temperature 0 to 55 °C
Absolute accuracy (log)
± 0.04 dB
± 0.08 dB
Absolute accuracy (linear)
± 0.8%
± 1.7%
Relative accuracy (log)
± 0.08 dB
± 0.16 dB
Relative accuracy (linear)
± 1.6%
± 3.4%
Time base accuracy
0.01%
1 mW power reference
Power output 1.00 mW (0.0 dBm). Factory set to ± 0.5% traceable to the National Physical
Laboratories (NPL), UK
2
Accuracy For two years
± 0.5% (23 ± 3 °C)
± 0.6% (25 ± 10 °C)
± 0.9% (0 to 55 °C)
Frequency
50 MHz nominal
SWR
1.06 maximum (1.08 maximum for Option E41xA-003)
Connector type
Type N (f), 50 ohms
1 Power meter is within ± 5 °C of its calibration temperature.
2 National metrology institutes of member states of the Metre Convention, such as the National Institute of Standards
and Technology in the USA, are signatories to the ComitÈ International des Poids et Mesures Mutual Recognition
Arrangement. Further information is available from the Bureau International des Poids et Mesures, at
http://www.bipm.fr/

Find us at www.keysight.com Page 5
Table 2. Measurement speed for different sensor types.
Channel functions
A, B, A/B, B/A, A-B, B-A and Relative
Storage registers
10 instrument states can be saved via the Save/Recall menu
Predefined setups For common wireless standards (GSM900, EDGE, NADC, iDEN, Bluetooth
®
, IS-95
CDMA, W-CDMA and cdma2000
®
), predefined setups are provided
Trigger
Sources
Internal, External TTL, GPIB, RS232/422
Time resolution
50 ns
Delay range
± 1.0 s
Delay resolution
50 ns for delays < ± 50 ms; otherwise 200 ns
Hold-off
– Range
– Resolution
1 µs to 400 ms
1% of selected value (minimum of 100 ns)
Internal Trigger
Range
–20 to +20 dBm
Level accuracy
± 0.5 dB
Resolution
0.1 dB
Latency
500 ns ± 100 ns
Latency is defined as the delay between the applied RF crossing the trigger level and the meter
switching into the triggered state.
External trigger range High > 2.0 V, Low < 0.8 V; BNC connector; rising or falling edge triggered; input
impedance > 1 kW
Trigger out Output provides TTL compatible levels (high > 2.4 V, low < 0.4 V) and uses a BNC
connector
1 Fast speed is not available for 8480 and N8480 series sensors.
2 Maximum measurement speed is obtained by using binary output in free run trigger.
3 For E9320 sensors, maximum speed is achieved using binary output in free run acquisition.
Sensor type Measurement speed (readings/second)
Normal x 2
Fast
1, 2
E-Series E9320 peak and average sensors
Average only mode
20
40
400
Normal mode
3
20
40
1000
E-Series CW and E9300 average power sensors
20
40
400
8480 and N8480 Series sensor
20
40
NA

Find us at www.keysight.com Page 6
Sampling Characteristics
Sampling rate
20 Msamples/second
Sampling technique
Continuous sampling
Rear Panel Inputs/Outputs
Recorder output(s) Analog 0 to 1 V, 1 kW output impedance, BNC connector. Two outputs are
available on E4417A (channels A and B).
Remote input/output
– TTL output
Used to signal when measurement has exceeded a defined limit
– TTL input
Initiates zero and calibration cycle
Connector type
RJ-45 series shielded modular jack assembly
– TTL output High = 4.8 V max;
Low = 0.2 V max.
– TTL input High = 3.5 V min, 5 V max;
Low = 1 V max, –0.3 V min
RS-232/422 interface Serial interface for communication with an external controller. Male plug 9-pin
D-subminiature connector.
Trigger in
Accepts a TTL signal for initiating measurements, BNC connector
Trigger out
Outputs a TTL signal for synchronizing with external equipment, BNC connector
Ground
Binding post accepts 4 mm plug or bare wire connection
Line power
Input voltage range
85 to 264 Vac, automatic selection
Input frequency range
47 to 440 Hz
Power requirement
Approximately 50 VA (14 Watts)
Remote Programming
Interface GPIB interface operates to IEEE 488.2 and IEC-625. RS-232 and RS-422 serial
interfaces supplied as standard
Command language
SCPI standard interface commands
GPIB compatibility
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0.
Environmental Specifications
Operating environment
Temperature
0 to 55 °C
Maximum humidity
95% at 40 °C, (non-condensing)
Maximum altitude
3,000 meters (9,840 feet)