U-I-HPP-IPP_Kurzanleitung_EN-DE-FR-ES-PT-IT-NL_D39872.pdf - 第4页

4 GmbH+Co.KG D-91126 Schwabach FRG Äußere Rittersbacher Str . 38 Made in Germany memmert Schutzart DIN EN 60529 - IP 20 1 2 3 4 5 6 10 9 8 7 Typ: IN260 230 V~ 50/60 Hz DIN12880-2007-Kl.: 1 F-Nr.: B618.3008 7,4 A 1700 W N…

100%1 / 44
3
Behaviour in case of malfunctions and irregularities

irregularities,malfunctions or damage, immediately take the appliance out of service
and inform your superior.
Switching off the appliance in an emergency
Push the On/Off switch on the control panel and
disconnect power plug. This disconnects the ap-
pliance from the power supply at all poles.
Applied directives and standards
INx IFx UNx UFx IPPx HPPx
X 
Directive of the council for harmonisation of the laws of the member states on elec-
tromagnetic compatibility. Standards complied with:
DIN EN 61326-1:2013-07 EN 61326-1:2013
X 
Directive of the council on harmonisation of the laws of member states relating to
electrical equipment designed for use within certain voltage limits. Standards com-
plied with:
DIN EN 61010-1:2011-07 EN 61010-1:2010
DIN EN 61010-2-010:2015-05 EN 61010-2-010:2014
UNxxm UFxxm UNxxmplus UFxxmplus INxxm IFxxm
X 
Class I, According to directive 93/42/EEC, annex IX Rule 1 / Rule 12
INxxmplus IFxxmplus
X 
Class I, According to directive 93/42/EEC, annex IX Rule 1 / Rule 2 / Rule 12


UNxx - UFxx
UNxxplus - UFxxplus
INxx - IFxx
INxxplus - IFxxplus
UNxxm, UFxxm
UNxxmplus
UFxxmplus
INxxm, IFxxm
INxxmplus
IFxxmplus
IFxxbw
IPPxx
IPPxxeco
HPPxx
HPPxxeco
IPPxxplus
IPPxxeco plus
This appliance is exclusively intended for heating up
non-explosive substances and objects. Any other use is
improper and may result in hazards and damage.
The appliance may be used for heating fango, silicate
and APS packs for physical therapy and keeping them
warm.
The appliance may be used for temperature control of
rinsing and infusion solutions and contrast agents.
The appliance is used to heat non-sterile cloths and
covers.
Peltier-cooled incubators IPP, IPPeco are intended for
the storage of substances and samples, for determi-
nation of life expectancy as well as for cultivation and

      
incubators IPPplus, IPPeco plus may be used exclusi-
vely for temperature and climate testing of materials
and substances in the context of the procedures and

Improper use
Any other use is improper and may result in hazards and damage. The appliance is
not explosion-proof (does not comply with the German occupational health and safety
regulation VBG 24). The appliance may only be loaded with materials and substances
which cannot form any toxic or explosive vapours at the set temperature and which
cannot explode, burst or ignite. The appliance may not be used for the drying, evapora-
tion and burning-in of paints or similar materials, the solvents of which could form an
explosive mixture when combined with air. If there is any doubt as to the composition
of materials, they must not be loaded into the appliance. Potentially explosive gas-air
mixtures must not form, neither in the working chamber nor in the direct vicinity of
the appliance.
GENERAL INSTRUCTIONS FOR USE
Intended use
4
GmbH+Co.KG D-91126 Schwabach FRG
Äußere
Rittersbacher Str. 38 Made in Germany
memmert
Schutzart DIN EN 60529 - IP 20
1
2
3
4
5
6
10
9
8
7
Typ: IN260
230 V~ 50/60 Hz
DIN12880-2007-Kl.: 1
F-Nr.: B618.3008
7,4 A 1700 W
Nenntemp.: 300 °C
CONSTRUCTION AND DESCRIPTION
memmert
1
2
3
4
5
6
Appliance overview
Designation (nameplate)
Connections and interfaces
1
2 On/Off switch
3 Door handle
1 Type designation
2 Operating voltage
3 Applied standard
4 Protection type
5
1
2
3
4 Nameplate (Door frame inside)
5
6 USB interface (TwinDisplay)
6 Address of manufacturer
7 Disposal note
8 Temperature range
9
10 Appliance number
1
2
3
4
4 Power connection
5
DELIVERY, TRANSPORT AND SETTING UP
CAUTION
≤ 260 L > 260 L
You could be injured due to the weight of the device if you try to lift
it alone. Observe the required number of people to carry equipment.
Larger equipment must not be carried, but only transported with a lift
truck or forklift.
≤ 75 L
Setting up
Due to its centre of gravity, the appliance can fall over to the front and
injure you or other people. Always attach the appliance to a wall with
the tilt protection.
WARNING
The distance between the wall and the rear of the appliance must be at least 15 cm.
The clearance from the ceiling must not be less than 20 cm and the side clearance

in the vicinity of the appliance must be guaranteed at all times.
Putting into operation

reached the steady state.
-
nection and power ratings (see nameplate and Technical data). Make sure to establish

Lay the power cable so that
X no one can trip over it;
X it does not come into contact with any hot parts;
X it is always accessible and within reach so it can be disconnected quickly in the
event of failure or emergencies.
≥ 15 cm
You may get your hands or feet squashed when transporting and in-
stalling the appliance. Wear protective gloves and safety boots. When
grasping the bottom of the appliance, grasp it only on the sides.
CAUTION
The appliance could fall over and seriously injure you. Never tilt the ap-
pliance and transport it in upright position and without load only (except
for standard accessories such as steel grids or shelves). Appliances
with castors always have to be moved by two people.
CAUTION
NOTICE
  
at high humidity conditions, the appliance shall be stored unpacked at
normal conditions for at least 24 hours. During this period of time the
appliance shall not be connected to the power supply.
WARNING