Vision XP+ V AC / XP+ / XP / XS Page 35 2 Mainte nance 2.9 Co ndensate T rap Service Instructions V ersion 1.3 2.9.3 Cleaning the Catch P ans (B) Fig. 2-46 Pulling Out the Catch Pan The cat ch pans are lo cated in the bo…

Page 34 Vision XP+ VAC / XP+ / XP / XS
2 Maintenance
2.9 Condensate Trap
Service Instructions
Version 1.3
2.9.2 Clean the suction ducts and the Gauze Filter (A)
Fig. 2-44 Pulling Out the Gauze Filter
Fig. 2-45 Pulling out of the Gauze Filter
Wipe out the suction ducts behind
the fan wheels with the oven
cleaner.
The Gauze Filter are located in the
suction duct on the right side of the
Condensation Trap (see Fig. 2-22,
point A).
Consumable materials, tools:
• Oven cleaner
• Rags
• Rinsing bath
Procedure:
1. Pull the gauze filter out and set
it into a rinsing bath.
2. Clean the opening for the gauze
filter next to the condensate
trap with oven cleaner and rags.
3. Reinsert the cleaned gauze
filter.

Vision XP+ VAC / XP+ / XP / XS Page 35
2 Maintenance
2.9 Condensate Trap
Service Instructions
Version 1.3
2.9.3 Cleaning the Catch Pans (B)
Fig. 2-46 Pulling Out the Catch Pan
The catch pans are located in the
bottom row of the condensate trap
(see item B in Fig. 2-46).
Consumable materials, tools:
• Oven cleaner
• Rags
• Rinsing bath
Procedure:
1. Open both locking levers and
pull out the cartridge.
2. Pull out the catch pan, empty it
an set it into the rinsing bath.
3. Clean the opening for the catch
pan and the door at the
condensate trap with oven
cleaner and rags.
4. Reinstall the cleaned catch pan
and slide the cartridge back into
place.
5. Hook the water connections
back up again.

Page 36 Vision XP+ VAC / XP+ / XP / XS
2 Maintenance
2.9 Condensate Trap
Service Instructions
Version 1.3
2.9.4 Cleaning Work at the Cooler Units (C)
Fig. 2-47 Pulling Out the Cooler Unit
Fig. 2-48 Pulling Out the Filter
Fig. 2-49 catch pan
The cooler units are located in the
bottom row of the condensate trap
(see item C in Fig. 2-47).
Consumable materials, tools:
• Oven cleaner
• Rags
• Rinsing bath
Procedure:
1. Open both locking levers and
pull out the cooler unit (see Fig.
2-48).
2. Pull the filter out of the cooler
unit and set it into the rinsing
bath.
3. Lift the cooler out of the cooler
unit and set it into the rinsing
bath as well.
4. Pull out the catch pan, empty it
an set it into the rinsing bath.
Caution!
The seals must protrude from
the bath. Otherwise they will be
damaged.
5. Clean the seals with industrial
alcohol and rags.
6. Clean the opening for the cooler
unit at the condensate trap with
oven cleaner and rags.
7. Reassemble cleaned parts and
insert them into the condensate
trap.
Note!
The quick couplings should be
lubricated occasionally, using
suitable lubricant, thereafter audibly
click the couplings back in.