Service Manual WPC5 / WPC6
Page 4-128
4.2.4 Calibrating the Limit Switch
The plus and minus position of the limit switches need to be calibrated.
Calibrating the limit switch -
Select the feed axis (1).
Select the function
Calibrate limit switch -
(2).
Select
Next
(3).
Follow the instruction shown and continue with
Commit
(1).
Service Manual WPC5 / WPC6
Page 4-129
Confirm the new position at which the limit switch triggers with
Finish
(1).
Calibrating the limit switch +
Select the feed axis (1).
Select the function
Calibrate limit switch +
and continue with the same procedure as that
used for calibrating the limit switch.
Service Manual WPC5 / WPC6
Page 4-130
4.2.5 Teaching the WPTC Transfer Position
Select the feed axis (1).
Select the function
Teach WPTC alignment position
(2).
Select
Next
(3).
Insert the calibration pin Adjusting gauge small WPC [03052363-02] on the front end of the
feed axis (1).
Adjusting gauge small WPC
[03052363-02]