Alltest-Agilent-Keysight-E9902G-Datasheet-29440-.pdf.pdf - 第8页

P a g e F i n d u s a t w w w . k e y s i g h t . c o m P a g e 7 High Guard Ratio Tests Range Measurement Type Accuracy 10 k  6-wire (using ED, EN, & OXT opti ons) (Source Voltage: 1.0 VDC) (Guard Ratio: 1000 to 1 …

100%1 / 22
P
a
g
e
Find us at www.keysight.com Page 6
Unpowered / Analog In-Circuit Test Measurement Specification
Shorts and Opens
Source Impedance: 100
Parameter
Specification
Programmable Threshold Range
Short
21000
Open
21000
Programming Resolution
1.0
Accuracy
±(0.25% + 2.2 )
Programmable Settling Time
Minimum
0 µs
Maximum
3.2768 s
Default
50 µs
Programming Resolution
50 µs
Test Voltage
0.1 V DC
Resistor / Potentiometer
Unguarded Tests
Source Voltage: 0.1 VDC
Range
Measurement Type
Accuracy
0.1 10
4-wire (using ED, EN, & OXT options)
±1.5%
10 300
4-wire (using ED, EN, & OXT options)
±1.0%
300 10 k
4-wire (using ED, EN, & OXT options)
2-wire
±0.25%
±1.0%*
10 k 100 k
2-wire (using ED, EN, & OXT options)
2-wire
±0.25%
±1.0%
100 k 1 M
2-wire (using ED, EN, & OXT options)
2-wire
2-wire
±0.5% ±2.5%
1.0 M 10 M
2-wire (using ED, EN, & OXT options)
±5.0%
* Plus system residual 3.5
Unguarded Tests (Using DMC on ASRU-N)
Source Voltage: 0.1 VDC
Range
Measurement Type
Accuracy
1 300
4-wire (using ED option)
±1.0%
300 10 k
4-wire (using ED option)
2-wire
±0.25%
±1.0%*
10 k 100 k
2-wire (using ED option)
2-wire
±0.25%
±1.0%
100 k 1 M
2-wire (using ED option)
2-wire
±0.5%
±2.5%
1.0 M 10 M
2-wire (using ED option)
±5.0%
* Plus system residual 3.5
P
a
g
e
Find us at www.keysight.com Page 7
High Guard Ratio Tests
Range
Measurement Type
Accuracy
10 k
6-wire (using ED, EN, & OXT options)
(Source Voltage: 1.0 VDC)
(Guard Ratio: 1000 to 1 on both legs of guard circuit)
±2.5%
100 k
6-wire (using ED, EN, & OXT options)
(Source Voltage: 0.1 VDC)
(Guard Ratio: 1,000,000 to 1)
±1.0% (typ)
* High/Ultra-High guard ratio is not supported in tests using DMC on the ASRU-N card
Capacitor
Unguarded Tests
Source Voltage: 0.1 VAC
Dissipation Factor: 1.0
Range
Measurement Type
Accuracy
10 pF 0.5 µF
2-wire (using ED & OXT options)
±2.0%*
0.5 µF 10,000 µF
4-wire (using ED, EN, & OXT options)
±2.0%
* Plus system residual: ±1 pF with capacitor compensation, 0 to +40 pF typical without capacitor compensation.
Unguarded Tests (using DMC on ASRU-N)
Source Voltage: 0.1 VAC
Dissipation Factor: 1.0
Range
Measurement Type
Accuracy
10 pF 0.5 µF
2-wire (using ED option)
±2.0%*
0.5 µF 100 µF
4-wire (using ED option)
±2.0%
* Plus system residual: ±2.5 pF with capacitor compensation, 0 to +40 pF typical without capacitor compensation.
100k and 200k test frequencies are available when using DMC on the ASRU-N card.
Guarded Tests
Source Voltage: 0.1 VAC
Test Frequency: 1024 Hz
Dissipation Factor: 1.0
Guard Ratio: 1000 to 1 on both legs of guard circuit
Value
Measurement Type
Accuracy
1000 pF
6-wire (using ED, EN, & OXT options)
±6.0%
* Guard sense lines are not supported in tests using DMC on the ASRU-N card. Tests using source and detector sense lines are allowed.
Inductor
Unguarded Tests
Source Voltage: 0.1 VAC
Quality Factor: ω 1.0
Range
Measurement Type
Accuracy
5 µH 50 mH
4-wire (using EN option)
±2.0%*
50 mH 1.59 H
2-wire (using EN option)
±2.0%
1.59 H 10 H
2-wire (using ED and EN options)
±2.0%
10 H 100 H
2-wire (using ED and EN options)
±3.0%
* Plus system residual: 1 µH.
P
a
g
e
Find us at www.keysight.com Page 8
Unguarded Tests (using DMC on ASRU-N)
Source Voltage: 0.1 VAC
Quality Factor: ω 1.0
Range
Measurement Type
Accuracy
5 µH 50 mH
4-wire
±2.0%*
50 mH 1.59 H
2-wire
±2.0%
1.59 H 10 H
2-wire (using ED options)
±2.0%
10 H 100 H
2-wire (using ED options)
±3.0%
* Plus system residual: 1 µH. 100k and 200k test frequencies are available when using DMC on the ASRU-N card.
Guarded Tests
Source Voltage: 0.1 VAC
Test Frequency: 8192 Hz
Dissipation Factor: 1.64
Value
Measurement Type
Accuracy
10 mH
6-wire (using ED and EN options)
±5.0%
Guard sense lines are not supported in tests using DMC on the ASRU-N card. Tests using source and detector sense lines are allowed
Diode
Unguarded Tests
Default Current: 1 mA
Programmable Range: Up to 100 mA
Range
Measurement Type
Accuracy
± 0V 19 V
±(1.0% of reading + 4 mV)*
± 0V19 V
* Plus system residual: 3.5 mV/mA
Zener Diode
Unguarded Tests
Default Current: 1 mA
Programmable Range: Up to 100 mA
Range
Accuracy
± 0V 18 V
±(1.0% of reading + 4 mV)*
* Plus system residual: 3.5 mV/mA
High Voltage Zener
Default Current: 1 mA to 5 mA
Programmable Range: Up to 50 mA
Range
Accuracy
18V 60 V
± 3.0% of reading
* Test is executed in pre-powered mode.
Bipolar Transistor (Beta Test)
Unguarded Tests
Programmable DC Emitter Bias
Current Range: 100 µA to 100 mA
Beta Range
Accuracy
10 1000
±15%