ACT - Accurac y Check Tool / User Manual 09/2015 Edition 14 ► Apply the foil. ► Start at the wide sid e of the measurin g plate ( Fig ure 3-4 ). ► Smooth any air bubbl es out towards t he side. Figure 3-4: ACT measuring …

100%1 / 184
ACT - Accuracy Check Tool / User Manual 09/2015 Edition
13
NOTICE
Cleanliness of plate and glass components
To ensure proper measurement, make sure that the measuring plate and the
components are free of all contaminants (particularly the fiducials in the corners of the
glass components).
The glass components should be cleaned with glass cleaning agent (e.g. Viss universal
from Lever)
Clean the measuring plate and apply the double-sided adhesive foil as described below.
NOTICE
Applying the foil without air bubbles
Make sure that the foil is applied without air bubbles, particularly in the area around
the line markings on the measuring plate.
Applying the foil
Remove the narrow strip of protective paper from the foil (Figure 3-3).
Figure 3-3: Adhesive foil with protective paper
Key:
(1) Protective sheet on adhesive
foil
(2) Adhesive foil
ACT - Accuracy Check Tool / User Manual 09/2015 Edition
14
Apply the foil.
Start at the wide side of the measuring plate (Figure 3-4).
Smooth any air bubbles out towards the side.
Figure 3-4: ACT measuring plate with adhesive foil
Key:
(1) Protective sheet on
adhesive foil
(2) ACT measuring plate
Apply the foil as displayed in Figure 3-5.
Figure 3-5: ACT measuring plate with adhesive foil
Key:
(1) Protective sheet on
adhesive foil
(2) ACT measuring plate
ACT - Accuracy Check Tool / User Manual 09/2015 Edition
15
Take a cloth and smooth the foil beginning at the middle as displayed in Figure 3-6 and
Figure 3-7.
Figure 3-6: ACT measuring plate with adhesive foil
Figure 3-7: ACT measuring plate with adhesive foil - applying the foil