3.7.18 Lading flap close magnet
Spare Parts
Lading flap close magnet [03083123-xx]
Removal / Installation
The holding magnet is mounted on a pedestal (1) in the NSM panel with 3 "NB602 M4x8"
screws [03005643].
The 3 screws can be accessed from the outside through a hole in the underside of the panel
(2).
The magnet has a 20 cm long cable with a 2 pin plug firmly attached.
The magnet is plugged into connector X30a on the back of the "backplane control
module" via the "cable: holding magnet loading flap" [03083109-01].
The adjustment with respect to the distance of the magnet to the loading door via the
"magnetic plate", which is mounted on the tailgate.
1
2
Attention: Do not clamp the cable!
Lead the cable through the notch on
the magnet.
Service Manual WPC5 / WPC6
Page 4-111
4 Calibrating the Sensors in the Station Software
After replacing the sensors, you need to calibrate the new sensors. This can take several
minutes.
Before calibration, perform a complete reference run of the machine.
Before calibration, remove all waffle-pack tray carriers (WPTCs) from the WPC.
Log in as
Machine Service
.
Select the
Check sensors and functions
button.
Service Manual WPC5 / WPC6
Page 4-112
Select the required location (1). (location 1 in this example)
Select the
Advanced functions
button.
The relevant axis can be selected in the following view.
See:
"4.1 Calibrating the Lifting Axis"
[
4-114]
"4.2 Calibrating the Feed Axis" [
4-123]
"4.3 Calibrating the Load Axis (WPC6 only)" [
4-132]