Service Manua l WPC5 / WPC6  Page 4-123 4.2 Calibrating the Feed Axis Before calibration, rem ove all waffle -pack tra y carriers (W PTCs) from the W PC. Calibration of the feed axis consists of the follo wing functions…

100%1 / 143
Service Manual WPC5 / WPC6
Page 4-122
➢ Follow the instruction shown and confirm the changes made with
Finish
.
Service Manual WPC5 / WPC6
Page 4-123
4.2 Calibrating the Feed Axis
Before calibration, remove all waffle-pack tray carriers (WPTCs) from the WPC.
Calibration of the feed axis consists of the following functions
• "4.2.2 Teaching the Zero Point Position" [
➙
4-124]
• "4.2.3 Reference Proximity Switch (Bero)" [
➙
4-126]
• "4.2.4 Calibrating the Limit Switch" [
➙
4-128]
• "4.2.5 Teaching the WPTC Transfer Position" [
➙
4-130]
4.2.1 Requirements for Settings
Feed axis – order of settings
If the limit switches or sensors are replaced, you will need to recalibrate or reset these
accordingly.
A complete recalibration is only needed if the zero point position is changed.
The function for the
limit switch +
and
limit switch –
specified in the software is assumed by the
end position stoppers (bumpers) at the machine end and tower end.
Service Manual WPC5 / WPC6
Page 4-124
4.2.2 Teaching the Zero Point Position
The zero point position needs to be calibrated after the drive motor has been replaced.
➢ Remove all waffle-pack tray carriers (WPTC) from the WPC.
➢ Select the feed axis (1).
➢ Select the function
Teach zero position
(2).
➢ Select
Next
(3).
➢ Follow the instruction shown and continue with
Commit
(1).