88181945-01-06 Technical Reference Manual ( Production Iss 06 December 2020).pdf - 第18页

18 Issue 06 , December 2020 Empty Condensate Collector Condensate Collector Full (TCM Shut down) FIG 19 6. Fit drip tray plug to C o ndensate Container to prevent s p illage and allow C o ndensate Container to be carried…

100%1 / 42
17 Issue 06, December 2020
A sticker on the machine, adjacent to the Controller, can be used as a guide for the
procedure described (FIG 17).
FIG 17
4. The Process Temperature (PV) within the Printer should move towards and stabilize
within +/- 1°C of the Set Value Temperature (SV) in approximately 30 minutes (FIG18).
>40 min
<30 min
PV = SV +/- 1°C
PV ≠ SV +/- 1°C
ESCALATE
FIG 18
5. During TCM operation condensate will be produced by the Air Conditioning Unit (ACU).
Condensate will be collected in the Condensate Container. The beacon will indicate the
state of the Condensate Container. Failure to empty the Condensate Container will
result in automatic TCM shutdown to prevent further condensate being produced (FIG
19).
Monitor Condensate Level Regularly to Prevent Machine Shutdown.
18 Issue 06, December 2020
Empty Condensate
Collector
Condensate Collector
Full
(TCM Shut down)
FIG 19
6. Fit drip tray plug to Condensate Container to prevent spillage and allow Condensate
Container to be carried using the handle. Remove drip tray plug when replacing
container (FIG 20).
FIG 20
19 Issue 06, December 2020
If the Condensate Container is not replaced after emptying, or refitted with the
plug still fitted, condensate will collect on the lower tray. Sensor probes at the
front of the lower tray will detect the presence of condensate and cause the
TCM to shutdown (FIG 21).
Condensate Present on
Lower Tray
(TCM Shut down)
FIG 21
Remove the cap from the lower tray to allow condensate to drain from the tray and
empty any condensate within the cap to clear the fault. The TCM will automatically
commence operation (FIG 22).
FIG 22