Fault Indication – Heater Circuit Breaker Open (Automatic Shutdown) YES NO Isolate mai ns s upply from TCM, reset 10A Heater MCB a n d power up TCM. Is the Heater fun ct ioning correctly now? Is fault indica ti on still …

Warning Indication - Condensate Container
Over Full (Automatic Shutdown)
YES
NO
Has the Condensate Container been
emptied?
Remove connector CN8 from control PCB.
Has the fault Cleared?
Empty Condensate
Container
Replace Control PCB
Part No. 181984
YES
Replace Level
Sensor
Part No. 181995
NO
Use TCM CCT DWG 181948 as reference for further
fault finding.

Fault Indication – Heater Circuit Breaker Open
(Automatic Shutdown)
YES
NO
Isolate mains supply from TCM, reset 10A
Heater MCB and power up TCM. Is the
Heater functioning correctly now?
Is fault indication still present?
Replace Heater
Modules(s)
Part No. 173772
YES
Use TCM CCT DWG 181948 as reference for
further fault finding.
Check following components for failure:
Control PCB Part No.181984
10A MCB Part No. 181986
Contactor Part No. 181989.

No Beacon Lamps Illuminated
YES
NO
Is the green indicator lamp next to the isolator illuminated
when the isolator is turned on?
Is the 1A MCB in the ON position?
Restore mains
supply to TCM.
230V +/- 10% (207V
to 253V working
range) 50/60Hz
(If MCB is a ‘C’ Type then fit replacement ‘D’
Type Part No 202405).
Reset MCB does it stay ON?
YES
NO
NO
Is 24V DC present on PSU output
terminals?
NO
NO
Disconnect CN4 from Control PCB.
Reset MCB.
Does it stay ON?
YES
Replace PSU Part No. 181987
Reconnect CN4.
Check beacon lamps, replacement
Part No. 178761.
Or replace Beacon Part No. 178762
YES
Is there 24V DC across Pin 1 and Pins
2-6 on control PCB CN2?
Replace Control PCB
Part No. 181984
YES
NO
Disconnect CN4 from Control PCB.
Is 24V DC present on PSU output
terminals?
NO
Reconnect CN4
YES
Use TCM CCT DWG 181948 as reference for further
fault finding.