D39489-BA-PM-EN.pdf - 第23页
23 D39489 | Date 12/2019 Installation and connec tion Diaphragm pump PJ25481 - 950.50.1 8 12 Translation of origina l Installation Ins tructions , english , KNF 125530 - 31 6634 06/18 Protect the pump agains t grease a…

22
D39489 | Date 12/2019
Diaphragm pump PJ25481-950.50.18 Installation and connection
Translation of original Installation Instructions, english, KNF 125530-316634 06/18 11
6. Installation and connection
Only install and operate the pumps under the pneumatic operating
parameters and conditions described in Chapter 4, Technical Data.
Observe the safety precautions (see Chapter 3).
6.1. Installation of the pump
Choose a safe location (flat surface) for the pump.
Before installation, store the pump at the installation location to
bring it up to ambient temperature.
See Fig. 3 for mounting dimensions.
Fig. 3: Mounting dimensions PJ25481-950.50.18 (All dimensional toler-
ances conform to DIN ISO 2768-1, Tolerance Class V)
Install the pump so that the motor fan can intake sufficient
cooling air.
Make sure that the installation location is dry and the pump is
protected against rain, splash, hose and drip water as well as
other pollutions.
Make sure, that the installation location is accessible for
maintenance and service.
The IP protection class of the pump motor is indicated on the
type plate.
Install the pump at the highest point in the system to prevent
condensate from collecting in the pump head.
Protect the pump from dust.
Mounting dimensions
Cooling air supply
Installation location

23
D39489 | Date 12/2019
Installation and connection Diaphragm pump PJ25481-950.50.18
12 Translation of original Installation Instructions, english, KNF 125530-316634 06/18
Protect the pump against grease and oils.
Protect the pump from vibrations and jolts.
WARNING
Personal injury and/or damage to property because
of vibration
In conjunction with adjacent components, vibration of
the pump may result in crushing and/or damage to
these components.
Make sure that vibrations of the pump do not re-
sult in hazards associated with adjacent compo-
nents.
Protect the pump against contact and intrusion of foreign mat-
ter.
Foreign matter protection

24
D39489 | Date 12/2019
Diaphragm pump PJ25481-950.50.18 Installation and connection
Translation of original Installation Instructions, english, KNF 125530-316634 06/18 13
6.2. Electrical connection
DANGER
Extreme danger from electrical shock
Only have the pump connected by an authorized
specialist.
Only have the pump connected when the power
supply is disconnected.
When connecting the device to a power source, the relevant
standards, directives, regulations, and technical standards
must be observed.
Connecting pump
1. Compare the supply data with the data on the motor plate. For
maximum operating current of the pump see pump’s type
plate.
The voltage must not vary by more than + 10% and – 10%
from that shown on the type plate.
2. Insert the mains cable plug into a properly installed shockproof
socket.
Ensure sufficient dimensioning of the connection cables.
6.3. Pneumatic connection
CAUTION
Personal injury or damages to property by ejected
protective
plugs
If the
protective plug at the pressure side of the
pump hasn’t been removed, it could be ejected be-
cause of the overpressure during operation.
Remove the protective plug during the installa-
tion.
Only connect components to the pump which are designed for
the pneumatic data of the pump (see Chapter 4, Technical
Data).
If the pump s used as a vacuum pump, safely discharge the
pump exhaust at the pump’s pneumatic outlet.
Connected
components
Pump exhaust