Test Research, Inc . TR7600 SII I Series User Guide – Maintenance 15 X -ray Source Manual Warm-U p 1) Please f ollow the following steps to do manual war m-up: I. Double-click the mouse on t he main program icon on the…

Test Research, Inc.
14 TR7600 SIII Series User Guide – Maintenance
II. In the Target field, confirm the path of the folder where the main program is located.
Figure 22: Target
III. Copy and paste the new WarmUp.ini file to this folder and overwrite the old
WarmUp.ini file.
IV. If you double-click on the new WarmUp.ini file, you can see the new setup of voltage
and current for manual warm-up. (Note: Although you can see the voltage and
current setups of manual warm-up, you unable to see the auto warm-up since they
are controlled by the tube software.
Figure 23: WarmUp.ini
5) In order to prevent overload and extend the life of X-ray tube, it is necessary to do
regular manual warm-up mandatorily.

Test Research, Inc.
TR7600 SIII Series User Guide – Maintenance 15
X-ray Source Manual Warm-Up
1) Please follow the following steps to do manual warm-up:
I. Double-click the mouse on the main program icon on the desktop.
II. Click [Utility]-->[Tool]-->[X-Ray].
III. Click ["Start Warm Up Procedure"] and it will take about 20 minutes to finish the
warm-up procedure.
Figure 24: Manual warm up
2) Based on the overload frequency, different machines have different manual warm-up
requirements to maintain the X-ray tube.
a) For new machines, please follow the suggestions below to maintain X-ray tube.
Please do warm-up maintenance manually twice per week after machines are
delivered to your assembly lines.
Please do warm-up maintenance manually once every day after machines have
been delivered to your assembly lines for more than six months.
b) For machines with occasional overload, please not only report the overload voltage
to TRI, but also follow the suggestions below to maintain X-ray tube.
To report the overload voltage to TRI:
I. Double-click the mouse on the main program icon on the desktop.
II. Click [Program] -->[Set Scan Information].

Test Research, Inc.
16 TR7600 SIII Series User Guide – Maintenance
Figure 25: Set scan information
III. In the pop-up window, input Voltage=90Kv and current=200μA and then
click [Ok] to finish.
Figure 26: Set scan info Dialog
IV. Click [On] on the main screen to activate the X-ray tube manually. After
voltage and current have reached the set values, observe the screen to
see if overload happens for 2 seconds.
Figure 27: Warm up button