技术资料.pdf - 第4页
4 Optimized for Te st Syste ms The 3498 0A has th e perfor mance yo u need for mediu m - to hi gh - density s witchin g/measurem ent applicat ions such as des ign verificat ion, fu nctiona l test, and data ac quisitio n.…

3
High-performance Data Acquisition System
Provides a Low-cost, Highly Flexible
Measurement Platform
Suppose you use automated test equipment for design validation or manufacturing. In that case, you now
have a cost-effective solution for many test system applications. The 34980A multifunction
switch/measure unit provides functionality that is easy to set up and use with a fast startup time. The
34980A helps you lower your cost of tests and accelerate your test-system integration and development
The 34980A handles system switching up to 26.5 GHz and provides basic measurements and system
control. It also offers DMM measurements, counter/ totalizer functionality, digital I/O with pattern
capabilities, and analog outputs with basic waveforms— all in one low-cost, compact box. And with its
standard connectors and software drivers, computer-standard I/O, and Web browser interface, the
34980A easily integrates into electronic functional test and data acquisition systems.
Flexible Switching, Measurements, and
System Control
The 34980A accommodates up to 8 plug-in modules to give you the flexibility you need. Choose from 21
different modules to define your own configuration. You can buy what you need now and add to it or
reconfigure it as your requirements change.
Whether you are measuring temperature, AC or DC voltage, resistance, frequency, current, or custom
measurements, the 34980A offers the functionality you need in a single box. Switch in different
measurements with high-performance signal switching up to 300V with no external signal conditioning
required. Choose between different switch types and topologies with frequency ranges from DC to 26.5
GHz. The 34980A offers high-density multiplexers for scanning multiple channels, matrices for connecting
multiple points at one time, and general-purpose switches for simple control and high power needs.
Use the 34980A to individually route signals or monitor multiple signals over a specified period of time —
monitor a single channel or multiple channels, set alarms, and identify irregularities.
The 34980A offers flexible choices for system control. You can control external devices such as
microwave switches, attenuators, solenoids, and power relays. Or you can use the digital inputs to sense
limit-switch and digital-bus status.

4
Optimized for Test Systems
The 34980A has the performance you need for medium- to high-density switching/measurement
applications such as design verification, functional test, and data acquisition. Your signals are switched to
the right measurement device without compromising signal integrity. Switch your signals to the optional
internal DMM and achieve optimal throughput on switch closure time. Or, if you prefer, you can easily
connect to external instruments such as DMMs, scopes, power supplies, and more. What’s more, with the
built-in Ethernet interface, you can control the 34980A and collect data from anywhere on the network.
For a complete modular solution, use the 34980A together with PXI modular products when more
capability is required.
The rugged instrument comes with a variety of system-ready features:
•
Web browser interface shows settings at a glance and provides remote access and control when
using a browser that supports Java applets.
•
Self-guiding front panel to configure, troubleshoot or view data
•
Low EMI and efficient system cooling
•
Heavy-duty cabling and connection options
•
Flexible rack mounting options
•
Relay counters help predict end-of-life
•
In-rack calibration for reduced maintenance time
•
DMM measurement accuracies include the switch for simple calculations
Make system connections easily and quickly with simple, reliable connection options:
•
Built-in Ethernet, USB 2.0, and GPIB connectivity
•
Low-cost, standard 50- or 78-pin Dsub connectors and cables
•
Detachable terminal blocks with strain relief
•
Mass interconnect solutions
Keysight Technologies, Inc. IO Libraries Suite enables you to quickly establish an error-free connection
between your PC and instruments—regardless of vendor. The IO Libraries provide robust instrument
control and enable you to work with the software development environment you choose.
The 34980A provides four 2-wire internal analog buses for easy signal routing. You can route your
measurements directly to the internal DMM, or you can connect to external instruments through the
analog bus connector on the rear of the mainframe. And since you have four 2-wire buses, you can
dedicate one bus for use with the internal DMM and use the other three buses for module extensions or
additional signal routing between modules which can reduce wiring complexity.
You can define up to 500 switch sequences to control complex signal routing and the order of switch
closures. Assign a sequence, give it a name and then execute it with the name you create. Switch
sequences are downloaded and stored in the instrument for ease of programming and increased
throughput.
External trigger capabilities make it easy for you to time and synchronize measurements and other events
and control when to begin or end an acquisition.

5
Measurements You Can Trust
Get proven performance from Keysight instruments with the resolution, repeatability, speed, and accuracy
you’ve come to expect.
The 34980A offers built-in signal conditioning and modular flexibility. When you use it with the internal
DMM, you can configure each channel independently for the measurements you choose. It includes a
variety of features that give you confidence in your measurements:
•
6½ digits of resolution with .004% of accuracy with DC voltage measurements
•
Alarms per channel—high limit, low limit, or both
•
Math functions—use Mx+B for custom linear conversions and converting raw inputs
•
Built-in thermocouple reference for temperature measurements (34921T)
•
Time-stamped readings
•
Add more formulas with BenchVue DAQ
The integrated DMM is mounted inside the mainframe and does not consume any of the eight user-
available slots. You can access the DMM through any switch module that connects to the analog bus, or
directly from the analog bus connector on the rear of the mainframe. The internal DMM gives you the
flexibility to measure 11 types of inputs:
•
Temperature with thermocouples, RTDs, or thermistors (with 34921A)
•
DC and AC voltage
•
2- and 4-wire resistance
•
Frequency and period
•
DC and AC current
You can control the DMM directly or configure it to work in conjunction with the switches. Each switch
channel can be configured independently for measurement functions, scale factors, and alarm limits.
Advanced measurement features such as offset compensation, variable integration time, and delay are
also selectable on a per-channel basis.
The DMM inputs are shielded and optically isolated from the 34980A’s earth-referenced circuitry and
computer interface, and as a result, you get up to 300 V of input isolation. This is important for reducing
ground-loops and common-mode voltage errors associated with long wiring runs and floating sources.
Simple DMM calibration is accomplished with just the analog bus connection on the rear panel of the
mainframe. You don’t need to remove the mainframe from the rack or dedicate a channel for calibration.