i3070series6_Site_Preparation+11-12.3.pdf.pdf - 第19页
Structural Requirements 3 Site Preparation 3-3 Moving Access Requirements The system is shipped fr om the fa ctory on one large crate (on a pallet) and several smaller crates or boxes. If the large t est hea d crate is t…

3-2 Site Preparation
3 Structural Requirements
Floor Requirements
The i3070 testhead is the heaviest part of the system. The controller and test
development stations present no special load-bearing concern. The following table
lists the approximate weights of the systems.
i307x
Anti-Static Surface
Static electricity is destructive to your production process and to your board test
system. Careless handling and poor planning can cost you yield and system
reliability.
This is not an exhaustive description of anti-static precautions but a reminder as
you plan your system area:
• Anti-static flooring. Plan to use an anti-static floor covering or mats.
• Grounding straps. Plan for foot straps in conjunction with anti-static flooring
and wrist straps for system operators. The testhead has external connectors for
wrist straps.
• Anti-static DUT storage. Plan for anti-static tote bins for your
devices-under-test and storage for anti-static bags.
Table 3-1 System weights
System Type Product Weight
E9903G Testhead
523 kg (1153 lb)
Testhead on pallet
776 kg (1710 lb)
E9902G Testhead
425 kg (937 lb)
Testhead on pallet
678 kg (1495 lb)
E9905G Testhead
386 kg (851 lb)
Testhead on pallet
561 kg (1237 lb)

Structural Requirements 3
Site Preparation 3-3
Moving Access Requirements
The system is shipped from the factory on one large crate (on a pallet) and several
smaller crates or boxes. If the large testhead crate is too large to move to your
system’s destination, you will need to unpack the testhead and roll it on its casters.
Other smaller boxes, containing the controller and test development hardware,
should not present access problems.
This section contains:
• Dimensions of the Crated System
• Dimensions of the Uncrated System
• Access Requirements
Dimensions of the Crated System
Table 3-2 Dimensions of the crated system
System Type Dimension Measurement
E9903G/E9902G Width 2153 mm
Depth 1220 mm
Height 1400 mm
E9905G Width 1617 mm
Depth 1220 mm
Height 1400 mm

3-4 Site Preparation
3 Structural Requirements
Dimensions of the Uncrated System
When removed from the crate, the testhead can be rolled on its casters.
Table 3-3 lists the dimensions of the uncrated testhead.
Equipment Rating
• Pollution degree 2
• Installation category II
• Altitude: Up to 2000 m
• Humidity: 5% to 80%
• Electrical supply: 208 Vac or 415 Vac, 50/60Hz, 30 A max, 3 phase
• Temperature: 0°C to 40°C
• Mains supply voltage fluctuations are not to exceed 10% of the nominal supply
voltage
This equipment is for indoor use only.
Table 3-3 Dimensions of the uncrated system
System Type Dimension Measurement
E9903G/E9902G Width 1490 mm
Depth 935 mm
Height 887 mm
E9903G/E9902G with side pod Width 1765 mm
Depth 935 mm
Height 887 mm
E9905G Width 954 mm
Depth 935 mm
Height 887 mm
E9905G with side pod Width 1229 mm
Depth 935 mm
Height 887 mm