141-400-57-probestationaccessoriescatalog.pdf_-_141-400-probestationaccessoriescatalog_1.pdf.pdf - 第126页

PN 141-400 www.formfactor.com Probe Mounts/Holders and Pro bes • 1 10 Revision 57, February, 2023 Probe Station A ccessories RF A- MN5 G- EW -TH Feat ur es • Eas t/W est arm enables mounting of M aury Nano5G f requency e…

100%1 / 169
PN 141-400 www.formfactor.com Probe Mounts/Holders and Probes 109
Revision 57, February, 2023 Probe Station Accessories
RFA-F110-WH-TH
Features
West probe arm enables
mounting of Focus
Microwaves Delta Tuner
110 GHz, harmonic, (M-
1102XX / M-1105XX), source
Compatible with RF
MicroChamber TopHat
(includes boot/FlexShield) o r
open (non-To pHat)
configurations
Non-Tophat plates (for closer
coupling) and TopHat adapter included
Compatibility
RPP404, RPP404-W, RPP504 positioners
RFA PROBE ARMS, RPP404/RPP404-W /RPP504
(S
INGLE SWEEP BROADBAND MEASUREMENTS FROM DC
TO 220GHZ)
RFA-K220-E
Features
East probe arm enables mounting of
N5291/VDI with T-Wave BB probe
Compatible with open (non-TopHat)
configurations
Compatibility
RPP404, RPP404-W, RPP504
positioners
RFA-K220-W
Features
West p robe arm enables mounting of
N5291/VDI with T-Wave BB probe
Compatible with open (non-TopHat)
configurations
Compatibility
RPP404, RPP404-W, RPP504
positioners
RFA-K220-E-TH
Features
East probe arm enables mounting of
N5291/VDI with T-Wave BB probe
Compatible with RF MicroChamber
TopHat (includes boot/FlexShield) or
open (non-TopHat) configurations
Compatibility
RPP404, RPP404-W, RPP504
positioners
RFA-K220-W-TH
Features
West p robe arm enables mounting of
N5291/VDI with T-Wave BB probe
Compatible with RF MicroChamber
TopHat (includes boot/FlexShield) or
open (non-TopHat) configurations
Compatibility
RPP404, RPP404-W, RPP504
positioners
RFA PROBE ARMS, MAURY NANO5G
RFA-MN5G-EW
Features
East/West arm enables
mounting of Maury Nano5G
frequency extender
Compatible with open (non-
TopHat) configurations
Compatibility
RPP404, RPP404-W,
RPP504 positioners
RFA-F110-WH-TH with tuner
(positioner shown, not included)
Arm, positioner, and probe
assembly shown
(transparent parts not included)
PN 141-400 www.formfactor.com Probe Mounts/Holders and Probes 110
Revision 57, February, 2023 Probe Station Accessories
RFA-MN5G-EW-TH
Features
East/West arm enables
mounting of Maury
Nano5G frequency
extender
Compatible with RF
MicroChamber TopHat
(includes boot/
FlexShield) o r open
(non-To pHat)
configurations
Compatibility
RPP404, RPP404-W,
RPP504 positioners
RFA PROBE ARMS STORAGE POD
780-00169 RFA Series Probe Arms Storage Pod
Features
Ena bles safe storage of
populated RFA arms
Stores RFA probe arms while
configured with probes, VNA
frequency extenders and/or
tuners, and as sociated port
savers and flex-shields.
Ena bles safe and efficient RFA
arm changes and band
swapping
Reduces the risk of damage to
the probe d uring arm changes
Fitting hardware enables
placement of the VDI Mini
VNAX power supply to keep the
extender warm between band
swaps, minimizing drift and
maximizing prober utilization
Integrated trays enable storage of probing accessories
Ena bles probe assembly from a sitting position, away from the
prober
A hinged, an ti-static, cleanroom-safe window prot ects probes from
accidental damage and prevents accidental probe tip contact,
while allowing safe, ergonomic probe arm loading
Ergonomic hand grips ensure safe and secure handling of valuable
test equipment
Each pod stores o ne populated probe arm
Storage pod can be placed on the prober platen during band
swaps on the CM300 and Summit 12k / Summit 200 probe stations
(depending on configuration)
Specifications
Dimensions (L x W x H): 350 x 175 x175 mm (or larger, depending
on stored parts [e.g., RFA-K220])
Compatibility
(* Denotes all variants)
Compatible with, but not limited to, TopHat and non-TopHat
configurations with the following arms:
RFA-67* - Coaxial sub 67 GHz pr obes
RFA-K120* - Keysight Coaxial N5295AX03, used as part of
N5291A extenders up to 130 GHz
RFA-K220* - N5291/VDI with T- Wave BB probe
RFA-VDI* - VDI Waveguide Mini models and S geometry
Waveguide probes to 500 GHz, for both Tophat and non-Top Hat
RFA-F* - Focus delta tuner arms
Compatible with both FM4910 and UL94-V0 industry standards for
safety against combustion and smoke generation
TUNER INTEGRATION KITS
780-00612 Integration Kit for Focus Microwave Tuner
M-4030 or L-4030 on CM300
Features
Enables integration of
the Focus Microwave
Tuner M-4030 and L-
4030 on the CM300
Positioner and tuner are
supplied by Focus
Microwave
Specifications
eVue cover is required and included
Configuration
Qty o ne required for each side
Configurable with both East and West tuners
Compatibility
CM300, Non-TopHat solution only
Arm, positioner, and probe
assembly shown
(transparent parts not included)
PN 141-400 www.formfactor.com Probe Mounts/Holders and Probes 111
Revision 57, February, 2023 Probe Station Accessories
DC PROBE ARMS, MICROALIGN
138687 Probe Arm, DPP2xx/DPP3xx, Triax
Features
For use with MicroAlign systems
Highly isolated needle clamp for standard
single tips like PTT needles
Stainless steel design for h igh rigidity and
minimum thermal drift
Requires adaptation unit 138697 for use with
DPP3xx-S/DPP450-S
Specifications
0.75 m (2.5 ft) 50 Ohm low noise triax cable and 3 lug Triax plug
DC leakage <10 fA at ambient temperature
For use with temperature chucks up to 300°C (572°F)
Compatibility
PA300 MicroAlign + DPP2xx, DPP3xx-S, DPP450-S (and former
PH110, PH150, PH400)
138690 Probe Arm, DPP2xx/DPP3xx, Triax, Kelvin
1Tip
Features
For use with MicroAlign systems
For triaxial I/V-, Kelvin- and C/V (with CX-TRX
adapter) tests
Stainless steel design for h igh rigidity and
minimum thermal drift
Highly isolated needle clamp for single tips
like PTT needles (signal from force and sense already connected
inside chassis)
Connector sockets for CommonRing cables (C/V-Test, 1 cable
included), guard potentials from force and sense are connected to
these sockets
Requires adaptation unit 138697 for use with DPP3xx-S/DPP450-S
Specifications
2 x 0.75 m (2.5 ft) 50 Ohm low-noise triax cable and 3 lug Triax
plug
For use with temperature chucks up to 300°C (572°F)
Compatibility
PA300 MicroAlign + DPP2xx, DPP3xx-S, DPP450-S (and former
PH110, PH150, PH400)
138697 Adapter for Probe Arms DPP2xx/DPP3xx to
Positio ner DPP3xx
Features
Converts DPP3xx-S/DPP450-S front
plate for use with DPP2xx/DPP3xx
probe arms for DC and RF
applications on MicroAlign systems
Compatibility
MPS150, EPS150, CM300xi stations, PM/PA stations + DPP3xx-S,
DPP450-S (and former PH15 0)
138700 Probe Arm, DPP2xx/DPP3xx, Pico Probe
Features
For MicroAlign systems
Holding fixture for Active Pico probes from
GGB Pico probe 10…34 (Pico probe not
included)
Stainless steel design for high rigidity and
minimum thermal drift
Requires Test Equipment Interface 144987
for use at MicroAlign Adva nced systems
Requires adaptation unit 138697 for use with DPP3xx-S/DPP450-S
Compatibility
PA300 MicroAlign + DPP2xx, DPP3xx-S, DPP450-S (and former
PH110, PH150, PH400)
144866 Probe Arm, DPP2xx/DPP3xx for DCP Probe
Features
For MicroAlign systems
For triaxial I/V-, Kelvin-, C/V (with CX-TRX
adapter) and 1/f applications
Holding fixture for advanced DC coaxial
probes like AP&T Probes, DCP and
DCP-HTR probes
Stainless steel design for high rigidity and minimum thermal drift
Requires adaptation unit 138697 for use with DPP3xx-S/DPP450-S
Includes (2) 0.7 m (2.3 ft) cables, 3-lug Triax-SSMC for IV an d
Kelvin measurements
Specifications
DC leakage <10 fA at ambient temperature
Compatibility
PA300 MicroAlign + DPP2xx, DPP3xx-S, DPP450-S (and former
PH110, PH150, PH400)