Installation Manual, Station Software Version 605.03 SP1 07/2011 Edition
5.3.3 Calibrating the touch screen
5.3.3.1 Calibrating the touch screen for the X-series and CA-series
→ Choose Start Æ Settings Æ TouchWare, to start the TouchWare Properties program.
Fig. 5-2: TouchWare Properties
→ Open the Multiple Monitors tab.
Fig. 5-3: TouchWare Properties – Multiple Monitors tab
→ Click the Map Touch Screens button.
The "Multiple Monitor Layout" dialog box appears.
25 of 48
Installation Manual, Station Software Version 605.03 SP1 07/2011 Edition
Fig. 5-4: "Multiple Monitor Layout" dialog box
→ Click the Mirrored Monitors button.
→ Confirm your entry with OK.
The touch screens are now mapped and the "Multiple Monitors" tab is displayed again.
Fig. 5-5: TouchWare Properties – Multiple Monitors tab (example for Monitor 1)
26 of 48
Installation Manual, Station Software Version 605.03 SP1 07/2011 Edition
Fig. 5-6: TouchWare Properties – Multiple Monitors tab (example for Monitor 2)
→ Click the number of the monitor you wish to calibrate.
→ Open the Calibrate tab.
Fig. 5-7: TouchWare Properties – Calibrate tab (example for Monitor 1)
→ Click the Calibrate button. An icon representing the calibration is shown on screen. An icon
representing a hand guides you through the calibration steps by pointing to the calibration icon you
must touch.
Fig. 5-8: Calibration marks
27 of 48