00197787-02_SI_SIPLACE_HeadVerification_EN.pdf - 第57页

6 Description of the test results 6.12 Force measurement P&P module (TH) Software Manual SIPLACE Head Verification 03/2018 57 6.12.3 Interpretation of the results obtained ‘Force’ error Cause Solution Force sensor se…

100%1 / 90
6 Description of the test results
6.12 Force measurement P&P module (TH)
56 Software Manual SIPLACE Head Verification 03/2018
6.12 Force measurement P&P module (TH)
6.12.1 Measurement principle
During force measurement, the placement force is measured for the segment at several angles.
The following force measurements are available for P&P modules:
Force 1N (dynamic)
Force 2N (dynamic)
Force 5N (dynamic)
Force 10N (dynamic)
Force 15N (dynamic)
6.12.2 Measurement result
The following image shows an example of the Force2N(dynamic) measurement:
Fig.44: Result view – Force 2N (dynamic)
1 Table showing the Force measured at different angles.
2 Graph showing the Force measured at different angles and the limits specified.
3 Graph showing the Force measured at different angles in more detail.
4 Graph showing a sample of the Force measurement in detail.
6 Description of the test results
6.12 Force measurement P&P module (TH)
Software Manual SIPLACE Head Verification 03/2018 57
6.12.3 Interpretation of the results obtained
‘Force’ error
Cause Solution
Force sensor settings misaligned Adjust the spring force settings.
Force measuring unit defective Exchange the force measuring unit.
Exchange / repair the P&P module.
Trick & feather unit defective Exchange the Trick & feather unit.
Exchange / repair the P&P module.
6 Description of the test results
6.13 Force sensor calibration / Force sensor calibration (final)
58 Software Manual SIPLACE Head Verification 03/2018
6.13 Force sensor calibration / Force sensor calibration (final)
6.13.1 Measurement principle
The Force sensor calibration serves as a preparatory step to ensure the force sensor is ready for
the subsequent force measurement.
The Force sensor calibration (final) is performed after the endurance run to verify the system re-
liability after being used for several cycles.
During the Sensor calibration, the Z-axis is moved down onto the force sensor several times
at 8 different angles to check if the force sensor is ready to use.
During the Spring preload measurement, the Z-axis is moved down onto the force sensor
several times at 8 different angles to check the mechanical position of the force sensor actu-
ator.
The Force 1N, Force 15N and DP positioning time measurement is used as a quick check
to verify the Sensor calibration and Spring preload measurement.
6.13.2 Measurement result
Fig.45: Result view - Force sensor calibration
1 Values determined for the Sensor calibration and Spring preload measurement.
2 Values determined for Force 1N, Force 15N and DP positioning time during the quick
check.
3 Graphs showing the Sensor calibration values for the tolerances allowed and in detail.
4 Graphs showing the Spring preload values for the tolerances allowed and in detail.