Page 1-90 444C-E-MMA01-A09-05 Check Every 12000 Hours (Once T wo Y ears) 8. Connect the pipe and wiring as be recorded. 9. Set the head cover . 10. T urn ON the power . 11. Press Machine adjust . • The “ Machine adjustme…

Page 1-89
REGULAR
MAINTENANCE
1
444C-E-MMA01-A09-03
Check Every 12000 Hours (Once Two Years)
4. Remove the wiring connected to the
valve.
∗ Write down the setting position and wiring
name. The setting working will be smooth.
5. Remove the pipe (tube) connected to
the valve (manifold).
∗ Write down the setting position and order. The
setting work will be smooth.
6. Remove the 2 bolt setting the bulb
(manifold).
7. Replace it with a new valve, and fix it
by the valve.
To the next page
444C-500P
444C-501P
444C-502P

Page 1-90
444C-E-MMA01-A09-05
Check Every 12000 Hours (Once Two Years)
8. Connect the pipe and wiring as be
recorded.
9. Set the head cover.
10. Turn ON the power.
11. Press
Machine
adjust
.
• The “Machine adjustment menu” screen
appears.
12. Press Return to origin .
• Returning to origin is performed.
To the next page

Page 1-91
REGULAR
MAINTENANCE
1
444C-E-MMA01-A09-04
Check Every 560 Hours (Once a Month)
13. Press
UNLOCK
+
START
.
14. Press Output check .
•“Output check” screen is displayed.
15. Select the operation position (the
table with the transfer head of which
the vacuum pressure is to be mea-
sured).
To the next page