4 Camera Calibration in Stations from SW705.03
4.7 Report File from SW 705.03 onwards
72 Operating Manual / Bedienungsanleitung Camera Calibration FCCS digital Kamerakalibrierung FCCS digital 04/2020
Fig.38: Checking the camera for soiling
ü The user level is Advanced production.
Go to the Maintenance view.
Select the Verification section.
Click on the Check for camera soiling
button.
Fig.39: Checking the camera for soiling
ü The last result (if present) will be shown!
Go to the Maintenance view.
Select the camera to be measured.
Place the contrast standard in the cali-
bration pocket and make sure that the con-
trast standard is correctly positioned.
Click on the Check for camera soiling
button.
ð The result will be shown and can be
saved in a file on the station computer
with the Generate report function. This
same file is then saved via the LineDia-
gnostic software on the Siplace Pro
computer, in the MaintenanceReports
folder.
On the station computer: C:\Sirio\Work\MaintenanceReports
On the SIPLACE Pro computer: C:\MaintenanceReports
4.7 Report File from SW 705.03 onwards
NOTICE
From the above mentioned station software version onwards a report file will auto-
matically be generated after the calibration.
Example: Report_FCCS_SX2_00004133_2011-01-31_15-20-40 currently from
SW705-03.pdf
NOTICE
Example for a report file of a SIPLACE SX2 machine
The row numbers of the first table references the results of the relevant camera in the fol-
lowing tables.
4 Camera Calibration in Stations from SW705.03
4.7 Report File from SW 705.03 onwards
Operating Manual / Bedienungsanleitung Camera Calibration FCCS digital Kamerakalibrierung FCCS digital 04/2020 73
Fig.40: List of cameras
The list of cameras gives an overview
of the calibration states and the
proper storage of the calibration val-
ues.
Fig.41: Camera details and calibration history
The camera details show the mater-
ial number, the serial number and the
number of flashes, i.e. the number of
components that have been centered
optically.
The calibration history shows the
dates of the first and the last cali-
bration as well as the number of cali-
bration processes.
4 Camera Calibration in Stations from SW705.03
4.7 Report File from SW 705.03 onwards
74 Operating Manual / Bedienungsanleitung Camera Calibration FCCS digital Kamerakalibrierung FCCS digital 04/2020
Fig.42: Calibration results
Calibration results for each success-
fully calibrated camera.
The result column shows for each illu-
mination level the set illumination dur-
ation with respect to the white stand-
ard as reference and the camera
amplification (camera gain).
The evaluation of the results with re-
spect to threshold values depends on
the camera type. See also table in
section 3.5.
The difference column shows the de-
viation to the last calibration.
4.7.1 Report file from SW 712.xx
Excerpt from the report file:
The results list has been extended by
a column on the right, to show the
camera cleaning status.
The degree of soiling is given as a
percentage and the bars show the
soiling in 1/3 steps from green to yel-
low and red.
If the display is in the red area, you
urgently need to clean the camera or
replace it.