ThermoFlex-Manual.pdf - 第108页

Section 8 8-6 ThermoFlex Thermo Scientific Process Fluid Pressure (P1) T ransducer Calibration The T her moFlex has been designed to minimize the need for calibration. How ever , if calibration is desired or recommended b…

100%1 / 138
Section 8
ThermoFlex 8-5
Thermo Scientific
xx.x
xx.x
xx
xx
r1 H
r1 L
r1 H
r1 L
xx.x C
mode
rtd1
PrES1
Flo
r rtd
A in
Aout1
CAL
enter
enter
enter
enter
enter
enter
StorE
mode
indicates the controller display is alternating
between the two displays.
indicates press and hold.
Figure 8-3 Internal Temperature Sensor Calibration
Shaded displays only appear on
chillers equipped with that option.
If you have any questions please contact Thermo Fisher Scientic's Sales, Service
and Customer Support.
Save calibration
Do not save calibration
Section 8
8-6 ThermoFlex
Thermo Scientific
Process Fluid
Pressure (P1)
Transducer
Calibration
The ThermoFlex has been designed to minimize the need for calibration.
However, if calibration is desired or recommended by our Sales, Service
and Customer Support, please use the following procedure.
This procedure requires a running chiller, a calibrated reference pressure
gauge and an external ow control valve.
Connect a calibrated reference pressure gauge to the outlet line. Using an
external ow control valve, increase the pressure to a suitable high-end
calibration point by closing the valve. Ensure the pressure is stabilized.
To enter the Calibration Loop ensure the controller display is displaying the
process uid temperature, see the diagram on the next page. Press and hold
the and then press the key. The display will indicate CAL.
Press the key and the controller will display rtd1. Press
until the controller displays PrES1. Press and the controller will
ash between P1H and the pressure.
Use to adjust the rate to match the reference pressure gauge.
Press the key to accept the value.
Decrease the pressure to a suitable low-end calibration point (avoid a zero
pressure). Ensure the pressure is stable.
The controller will ash between P1L and the pressure. Use to
adjust the rate to match the reference pressure gauge.
Press the key and StorE is displayed, press to save both
values, press to not save them.
Note After pressing the button at the StorE prompt wait
several seconds before proceeding to ensure that a bad calibration message
(Er 16) does not appear. Premature use of the keypad after pressing
may cancel the bad calibration error message.
Section 8
ThermoFlex 8-7
Thermo Scientific
xx.x
xx.x
xx
xx
P1 H
P1 L
P1 H
P1 L
xx.x C
mode
rtd1
PrES1
Flo
r rtd
A in
Aout1
CAL
enter
enter
enter
enter
enter
enter
StorE
mode
If you have any questions please contact Thermo Fisher Scientic's Sales,
Service and Customer Support.
Figure 8-4 Pressure (P1) Calibration
indicates the controller display is alternating
between the two displays.
indicates press and hold.
Shaded displays only appear on
chillers equipped with that option.
Save calibration
Do not save calibration