Section 7 ThermoFlex 7-11 Thermo Scientific Chiller s huts do wn Ensure button wasn't accidently pressed. Ensure the optional GFCI breaker located on the rear of the c hiller is in the up position. F or Ther moFlex90…

Section 7
7-10 ThermoFlex
Thermo Scientific
Checklist
Chiller will not start
Check electrical connections.
For rst time use, please refer to the quick start instructions included
with your chiller or the copy in this manual. The manual's copy follows
the Table of Contents.
Check the controller for error codes, see Error Codes in this Section.
Ensure the optional GFCI breaker located on the rear of the chiller is
in the up position.
For ThermoFlex900 - 10000 chillers ensure the circuit protector is in the
on ( I ) position.
Make sure supply voltage is connected and matches the chiller's
nameplate rating ±10%
No display on controller or display is 8888
For ThermoFlex900 - 10000 recycle the circuit protector on the rear of
the chiller.
Clearing Error Codes
Note the code in case it clears before you are done troubleshooting.
If desired, silence the audible alarm by pressing the up or down arrow
key.
If the chiller shut down, the controller will continue to ash the error code.
Press enter to clear the display and silence any alarm. Refer to Error
Codes in this Section. Once the cause of the shut down is identied
and corrected, start the chiller. If the cause was not corrected the error
code will reappear.
If the chiller is still running, press enter to see if the code clears, a limit
may have been only temporarily exceeded. If the error code does not
clear press
until the display ashes between the error code and
the temperature and then press enter. If the code still does not clear
refer to Error Codes in this Section.

Section 7
ThermoFlex 7-11
Thermo Scientific
Chiller shuts down
Ensure button wasn't accidently pressed.
Ensure the optional GFCI breaker located on the rear of the chiller is
in the up position.
For ThermoFlex900 - 10000 chillers ensure the circuit protector is in
the on ( I ) position.
Check the controller for error codes, see Error Codes in this Section.
The chiller is designed to shut down if not properly primed.
If able, pre-ll the process uid lines.
After start up continue adding process uid until the chiller and
uid lines are full.
If you need to pause chiller priming turn the chiller off using the
power button
on the front panel.
Not completely lling the chiller and process uid lines
could damage the chiller’s pump.
Make sure supply voltage is connected and matches the chiller's
nameplate rating ±10%.
Restart the chiller.
Chiller vibration
The optional pressure relief valve setting may be the cause. Change the
pressure setting ± 5 psi to eliminate the vibration.
Inadequate pump pressure
Ensure any user installed in-line valves are in the desired position.
Ensure the chiller’s process uid outlet is connected to the application’s
uid inlet and not the application’s uid outlet, see Section 3.
Ensure all connections are secure and that the proper sealant/lubricant
for the tting material is used.
Keep the distance between the chiller and the instrument being cooled
as short as possible.
Ensure tubing is straight and without bends. If diameter reductions are
required, make them at the inlet and outlet of your application, not at
the chiller.
CAUTION
Section 7
7-12 ThermoFlex
Thermo Scientific
Chiller will not circulate process uid
Check the reservoir level. Fill, if necessary.
Ensure the reservoir bag lter is not clogged.
Check the application for restrictions in the cooling lines.
Chiller requires >3 PSIG application pressure drop. If a bypass valve
has been installed add some restriction to the bypass line.
The pump motor overloaded. The pump's internal overtemperature
overcurrent device will shut off the pump causing the ow to stop. This
can be caused by low uid, debris in system, operating chiller in a high
ambient temperature condition or excessively conned space. Allow
time for the motor to cool down.
Make sure supply voltage matches the chiller's nameplate rating ±10%.
Inadequate temperature control
Verify the setpoint.
If the chiller is over-cooling, recycle the power.
Make sure the condenser/air lter is free of dust and debris.
Check the uid concentration, see Section 3.
Ensure chiller installation complies with site requirements, see Section 3.
Make sure supply voltage matches chiller nameplate rating ±10%.
For ThermoFlex900 - 5000 global voltage chillers ensure the chiller is
properly congured, see Appendix B.
If the temperature continues to rise, make sure your application's heat
load does not exceed the rated specications.
Check for high thermal gradients (e.g., the application load is being
turned on and off or rapidly changing).
Verify/adjust controller PID values, see next page.
Please contact Thermo Fisher Scientic Sales Service and Customer
Support if you need any additional information, see this manual's inside
cover for contact instructions.