Laboratory Equipment Div. 25 Nimble Hill Road Newington, NH 03801 D of C 094026.1 Form 094233 DWG # 094026 REV 17 Apr 07 DECLARATION OF CONFORMITY Manufacturer: Thermo Fisher Scientific Address: 25 Nimble Hill Road Newin…
Thermo Scientific
D-10 ThermoFlex
Appendix D
Thermo Scientific
Set Special Commands
These commands are product specic.
Master Sends: lc a1 a2 8D nn d1 d2 d3 d4 d5 d6 cs
Slave Returns: lc a1 a2 8D nn d1 d2 d3 d4 d5 d6 cs
Byte Master Slave
d1 Command byte
d2 Entered Value MSB
d3 Entered Value
d4 Entered Value
d5 Entered Value
d6 Entered Value LSB
Command Master sends Description Slave returns
0x00 CA 00 01 8D 02 d1 d2 cs Set analog option CA 00 01 8D 03 00 d2 d3 cs
d1 = command byte = 00
d2 = analog option byte
0x80 CA 00 01 8D 01 80 cs Request PM status CA 00 01 8D 03 80 d2 d3 cs
Set analog option command
d2 analog option byte
b.6 - b.7 = unused b.4 - b.5 = DAC enable b.2 - b.3 = DAC out b.0 - b.1 = analog in
0 = voltage 0 = voltage 0 = voltage
1 = millivolt 1 = millivolt 1 = millivolt
2 = current 2 = current 2 = current
3 = no change 3 = no change 3 = no change
Eg. Command to enable DAC, set DAC out to Voltage and set Analog in to millivolt
Master sends Slave returns
CA 00 01 8D 02 00 11 5E CA 00 01 8D 02 00 11 5E
Eg. Command to set DAC out to current without changing DAC enable or analog in
Master sends Slave returns
CA 00 01 8D 02 00 3B 34 CA 00 01 8D 02 00 19 56

Laboratory Equipment Div.
25 Nimble Hill Road
Newington, NH 03801
D of C 094026.1 Form 094233
DWG # 094026 REV 17 Apr 07
DECLARATION OF CONFORMITY
Manufacturer: Thermo Fisher Scientific
Address: 25 Nimble Hill Road
Newington, NH 03801 USA
Products: Refrigerated chillers and heat exchangers. Year of inception 2008
We declare that the following products conform to the Directives and Standards listed below:
Unit has a 15 digit part number consisting of UU C VV PP c XXXXXXX defined as follows:
UU = Unit type can be:
10 = TF 900 11 = TF 1,400 12 = TF 2,500 13 = TF 3,500 14 = TF 5,000
15 = TF 7,500 16 = TF 10,000 17 = TF 15,000 18 = TF 20,000 19 = TF 24,000
C = Cooling type and Temperature Range and can be a 1-4 inclusive, where:
1 = Air Cooled Standard Temp ( 5-40°C ) 2 = Air Cooled Hi Temp ( 5-90°C )
3 = Water Cooled Standard Temp ( 5-40°C ) 4 = Water Cooled Hi Temp ( 5-90°C )
VV = Unit voltage rating:
UU = 10, 11, 12, 13 & 14 UU = 15, 16 & 17 UU = 18 & 19
10 = 115V, 60Hz 1Ph 17 = 200/208/230V, 60Hz 3Ph 17 = 208/230V, 60Hz 3Ph
100V, 50Hz 1Ph 200V, 50Hz 3Ph
11 = 100/115V, 60Hz 1Ph 18 = 400V, 50Hz 3Ph 18 = 400V, 50Hz 3Ph
100/115V, 50Hz 1Ph 20 = 200/208/230V, 60Hz 3Ph
12 = 208/230V, 60Hz 1Ph 200-230V, 50Hz 3Ph
200V, 50Hz 1Ph 21 = 460V, 60Hz 3Ph 21 = 460V, 60Hz 3Ph
16 = 230V, 50Hz 1Ph 400V, 50Hz 3Ph 400V, 50Hz 3Ph
20 = 200/208/230V, 60Hz 1Ph
200/230V, 50Hz 1Ph
PP = Pump type, can be 10 through 25 inclusive.
c = Unit controller type, can be any digit from 1-6, inclusive.
X = Any digit from 0-9, used as sequential numbering only.
Equipment Class: Measurement, control and laboratory
Directives and Standards:
2004/108/EC – Electromagnetic Compatibility ( EMC Directive ):
EN 61326-1: 2006 – Electrical equipment for measurement, control, and laboratory use – EMC requirements,
EMC Class A
2006/95/EC – Low Voltage Directive ( LVD ):
EN 61010-1: 2004 – Safety requirements for electrical equipment for measurement,
control, and laboratory use – general requirements.
EN 61010-2-010: 2003 Particular requirements for laboratory equipment for the heating of materials.
Additional EMC Evaluations with Certificates:
EN 61000-3-2: 2006 Harmonics
EN 61000-3-3: 2008 Flicker
Manufacturer’s Authorized Representative: Date:
__________________________ 15 September 2011
Robin Wiley Compliance Engineering
WARRANTY
Thermo Fisher Scientic warrants for 24 months (excluding MD 1/MD 2 Magnetic Drive and P 1/P 2
Positive Displacement pumps which are warranted for 12 months) from date of shipment the Thermo Scientic
ThermoFlex chiller according to the following terms.
Any part of the chiller manufactured or supplied by Thermo Fisher Scientic and found in the reasonable judgment
of Thermo Fisher to be defective in material or workmanship will be repaired at an authorized Thermo Fisher Repair
Depot without charge for parts or labor. The chiller, including any defective part must be returned to an authorized
Thermo Fisher Repair Depot within the warranty period. The expense of returning the chiller to the authorized
Thermo Fisher Repair Depot for warranty service will be paid for by the buyer. Our responsibility in respect to
warranty claims is limited to performing the required repairs or replacements, and no claim of breach of warranty
shall be cause for cancellation or recision of the contract of sales of any chiller. With respect to chillers that qualify
for eld service repairs, Thermo Fisher Scientic’s responsibility is limited to the component parts necessary for
the repair and the labor that is required on site to perform the repair. Any travel labor or mileage charges are the
nancial responsibility of the buyer.
The buyer shall be responsible for any evaluation or warranty service call (including labor charges) if no defects are
found with the Thermo Scientic product.
This warranty does not cover any chiller that has been subject to misuse, neglect, or accident. This warranty does
not apply to any damage to the chiller that is the result of improper installation or maintenance, or to any chiller
that has been operated or maintained in any way contrary to the operating or maintenance instructions specied in
this Instruction and Operation Manual. This warranty does not cover any chiller that has been altered or modied
so as to change its intended use.
In addition, this warranty does not extend to repairs made by the use of parts, accessories, or uids which are either
incompatible with the chiller or adversely affect its operation, performance, or durability.
Thermo Fisher Scientic reserves the right to change or improve the design of any chiller without assuming any
obligation to modify any chiller previously manufactured.
THE FOREGOING EXPRESS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING BUT NOT LIMITED TO WARRANTIES OR MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
OUR OBLIGATION UNDER THIS WARRANTY IS STRICTLY AND EXCLUSIVELY LIMITED TO THE
REPAIR OR REPLACEMENT OF DEFECTIVE COMPONENT PARTS AND Thermo Fisher Scientic DOES
NOT ASSUME OR AUTHORIZE ANYONE TO ASSUME FOR IT ANY OTHER OBLIGATION.
Thermo Fisher Scientic ASSUMES NO RESPONSIBILITY FOR INCIDENTAL, CONSEQUENTIAL, OR
OTHER DAMAGES INCLUDING, BUT NOT LIMITED TO LOSS OR DAMAGE TO PROPERTY, LOSS
OF PROFITS OR REVENUE, LOSS OF THE CHILLER, LOSS OF TIME, OR INCONVENIENCE.
This warranty applies to chillers sold in the United States. Any chillers sold elsewhere are warranted by the afliated
marketing company of Thermo Fisher Scientic. This warranty and all matters arising pursuant to it shall be governed
by the law of the State of New Hampshire, United States. All legal actions brought in relation hereto shall be led
in the appropriate state or federal courts in New Hampshire, unless waived by Thermo Fisher Scientic.