Section 8 ThermoFlex 8-11 Thermo Scientific CAUTION CAUTION Decommissioning/ Disposal Decommissioning prepares equipment for safe and secure transpor tation. Laborator y Grade Eth ylene glycol (EG) is poisonous and ammab…

100%1 / 138
Section 8
8-10 ThermoFlex
Thermo Scientific
enter
enter
SEtuP
mode modemode
diA9
mode
mode
SP
mode
Press within 10 seconds
FLtrS
StEPS
0
enter
voLtS
xx.x
Beep
SEr1
Clearing SEr1
Message
Press the down arrow (approximately 11 times)
until StEPS is displayed.
Press and hold enter and then press the up arrow,
the display will show 0.
Press the down arrow to return to the temperature
display.
With SEr1 ashing press and hold enter and then
press mode three times, the controller should beep.
Press the mode key until diA9 is displayed. Press
enter.
Section 8
ThermoFlex 8-11
Thermo Scientific
CAUTION
CAUTION
Decommissioning/
Disposal
Decommissioning prepares equipment for safe and secure transportation.
Laboratory Grade Ethylene glycol (EG) is poisonous and
ammable. Before disposing refer to the manufacturer’s most
current MSDS for handling precautions.
Decommissioning must be performed only by qualied dealer using
certied equipment. All prevailing regulations must be followed.
Consider decommissioning the chiller when:
It fails to maintain desired specications
It no longer meets safety standards
It is beyond repair for its age and worth
Refrigerant and compressor oil must be recovered from equipment before
disposal.
Note Keep in mind any impact your application may have had on the
chiller.
Direct questions about chiller decommissioning or disposal to our Sales,
Service and Customer Support.
Handling and disposal should be done in accordance with the
manufacturers specication and/or the MSDS for the material used.
WARNING
Shipment
Storage
Follow the manufacturer's MSDS instructions if decontamination is
required.
Transporting and/or storing the chiller in near or below freezing
temperatures requires draining, see Draining in this Section.
Store the chiller in the temperature range of -25°C to 60°C (with
packaging), and <80% relative humidity.
If the chiller is stored for more than 90 days it must be ushed with
clean uid before operating.