Section 4 Operation
The controller controls temperature using a Proportional-Integral-Derivative
(PID) algorithm. It is designed with an easy to use operator interface.
This key is used to start and stop the chiller.
This key is used to navigate through the controller displays, to
make changes and to save changes once they are made. It is
also used to clear error codes.
This key is also used to navigate through controller displays.
The up arrow key is used to navigate through the controller
displays and to increase adjustable values.
The down arrow key is used to navigate through the
controller displays and to decrease adjustable values.
Figure 4-1 Basic Controller
mode
enter
+
-
enter
mode
+
-
Section 4
4-2 ThermoFlex
Thermo Scientific
Start Up
Setup
Note For rst time use, please refer to the quick start instructions included
with your chiller or the copy in this manual. The manual's version follows
the Table of Contents.
Before starting the chiller, double check all electrical and plumbing
connections. Have extra recirculating uid on hand. If the chiller will not
start refer to Section 7 Troubleshooting.
If the chiller is equipped with a deionization lter cartridge refer to Section
5 for installation.
• Place the optional GFCI breaker located on the rear to the up position.
• For ThermoFlex900 through 10000s, place the circuit protector located
on the rear to the on ( I ) position. The display will indicate a series of
upward scrolling bars (
).
• For ThermoFlex15000 and 24000s, the display will indicate a series of
upward scrolling bars (
) as soon as power is supplied.
• The bars will scroll upward indicating the controller is initializing. The
initialization takes approximately 15 seconds.
• When the bars disappear the controller display will go blank.
• Press the
key on the controller. The controller will show the process
uid temperature. The pump and refrigeration system will also start.
Note You can press the
key anytime after placing the circuit
protector to the on position.
If the auto restart is enabled and the chiller shuts down as a result of
a power failure, when power is restored the chiller will automatically
restart. Auto restart is enabled using the Setup Loop, see Setup Loop
in this Section.
Note After initial start up, check your plumbing for leaks.