Section 2
2-20 ThermoFlex
Thermo Scientific
ThermoFlex 3-1
Thermo Scientific
Section 3 Installation
Ambient Temperature Range* 10°C to 40°C (50°F to 104°F)
Relative Humidity Range 10% to 80% (non-condensing)
Operating Altitude* Sea Level to 8000 feet (2438 meters)
Overvoltage Category II
Pollution Degree 2
Degree of Protection IP 20
Never place the chiller in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Note Refer to the nameplate information on the rear of the chiller.
Air-cooled chillers retain their full rated capacity at 20°C setpoint in ambient
temperatures up to 25°C (77°F). For ambient temperatures above 25°C please
de-rate the cooling capacity 3% for every 1°C above 25°C (77°F), up to a
maximum ambient temperature of 40°C (104°F). Note that when operating
at a process temperature lower than 20°C the de-rate percentage may increase
due to additional gains from losses to ambient.
Note Depending on the setpoint and ambient temperatures, there may be a
heat gain or loss through the plumbing resulting in a variation from setpoint
temperature at the application inlet. Applications with large temperature
variations between ambient and setpoint temperatures, and/or long plumbing
lengths, may require additional insulation.
ThermoFlex2500 air-cooled chillers have a two-speed fan. Should the chiller's
internal ambient temperature reach 50°C for 30 seconds, or reach 53°C, the
fan speed will switch from slow speed to high speed to maintain internal
temperatures within acceptable limits. When the temperature reaches 44°C or
below for at least 15 minutes the speed will return to low. When in high speed

Note High speed is required for the chiller to achieve its 2500 watt cooling
capacity. At high-end operating conditions the fan can be set to run at high
speed all the time using the controller's Setup Loop, see Section 4.
Site Requirements
*Because of the decrease in air density, maximum temperature for the air entering an air-cooled
ThermoFlex is reduced by 1°C per 1,000 feet above sea level. In addition, cooling capacity is
reduced 1.2% per 1,000 feet above sea level.
CAUTION
Section 3
3-2 ThermoFlex
Thermo Scientific

drain back into the chiller and cause spillage. Thermo Fisher offers an
anti-drainback kit to prevent any spillage, see Section 5.
Air-cooled chillers can be installed with both sides blocked, or one side and
the rear. See Figure 3-1. The front of the chiller needs a minimum clearance
of 24". Air will enter the front of the system and exit through the sides and
rear.
Having two sides blocked can impact the chiller's performance due to
 
ensure that the following requirements are met:
Process Setpoint Temperature: Below 30°C (86°F)
Ambient: Below 40°C (104°F)
Before operating the chiller in conditions outside any of those listed on this

Support to review your installation.
Figure 3-1 Minimum Clearance