ThermoFlex-Manual.pdf - 第41页

Section 3 ThermoFlex 3-3 Thermo Scientific Electrical Requirements The chiller's constr uction pr ovides protection against the risk of electrical shock by gr ounding appropriate metal par ts. The protection will not…

100%1 / 138
Section 3
3-2 ThermoFlex
Thermo Scientific

drain back into the chiller and cause spillage. Thermo Fisher offers an
anti-drainback kit to prevent any spillage, see Section 5.
Air-cooled chillers can be installed with both sides blocked, or one side and
the rear. See Figure 3-1. The front of the chiller needs a minimum clearance
of 24". Air will enter the front of the system and exit through the sides and
rear.
Having two sides blocked can impact the chiller's performance due to
 
ensure that the following requirements are met:
Process Setpoint Temperature: Below 30°C (86°F)
Ambient: Below 40°C (104°F)
Before operating the chiller in conditions outside any of those listed on this

Support to review your installation.
Figure 3-1 Minimum Clearance
Section 3
ThermoFlex 3-3
Thermo Scientific
Electrical
Requirements
The chiller's construction provides protection against the risk of
electrical shock by grounding appropriate metal parts. The protection
will not function unless the power cord is connected to a properly
grounded outlet. It is the user's responsibility to assure a proper
ground connection is provided.
The chiller must be installed in accordance with the National Electrical Code
and the with reference to the information on the chiller's nameplate located
on the rear.
Locate the chiller so it is near, and has easy access to, its disconnecting device.
The user is responsible to ensure that the line cord provided meets local
 
The chiller is intended for use on a dedicated outlet. The ThermoFlex has
an internal circuit protection that is equivalent (approximately) to the branch
circuit rating. This is to protect the ThermoFlex, and is not intended as a
substitute for branch circuit protection.
DANGER
* Refer to Appendix A for country specific ratings. Continued on next page.
Electrical Service Requirements (Standard chillers):
ThermoFlex900 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
100 VAC 50 Hz 15A 5-15P
115 VAC 60 Hz 15A 5-15P
200 VAC 50 Hz 15A 6-15P
208-230 VAC 60 Hz 15A 6-15P
230 VAC 50 Hz *16A
1
, 15A
2
, 13A
3
-
ThermoFlex1400 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
100 VAC 50 Hz 20A 5-20P
115 VAC 60 Hz 20A 5-20P
200 VAC 50 Hz 15A 6-15P
208-230 VAC 60 Hz 15A 6-15P
230 VAC 50 Hz *16A
1
, 15A
2
, 13A
3
-
ThermoFlex2500 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
200 VAC P 1, P 2 Pump 50 Hz 15A 6-15P
208-230 VAC P 1, P 2 Pump 60 Hz 15A 6-15P
200 VAC T 1 Pump 50 Hz 20A 6-20P
208-230 VAC T 1 Pump 60 Hz 20A 6-20P
230 VAC 50 Hz *16A
1
, 15A
2
, 13A
3
-
Section 3
3-4 ThermoFlex
Thermo Scientific
Continued on next page.
Electrical Service Requirements (Standard chillers):
ThermoFlex3500/5000 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
200 VAC P 1, P 2 Pump 50 Hz 15A 6-15P
200 VAC T 1, P 3 Pump 50 Hz 20A 6-20P
200 VAC P 4 Pump 50 Hz 30A 6-30P
208-230 VAC P 1, P 2 Pump 60 Hz 15A 6-15P
208-230 VAC T 1, P 3 Pump 60 Hz 20A 6-20P
208-230 VAC P 4 Pump 60 Hz 30A 6-30P
230 VAC P 1 - P 4 Pump 50 Hz *16A
1
, 15A
2
, 13A
3
-
ThermoFlex7500/10000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
200 VAC P 2 Pump 50 Hz 16.5 30 L15-20P
200 VAC P 3 Pump 50 Hz 18.7 30 L15-20P
200 VAC P 5 Pump 50 Hz 22.3 35 L15-30P
200 VAC T 5 Pump 50 Hz 17.3 30 L15-30P
208-230 VAC P 2 Pump 60 Hz 16.5 30 L15-20P
208-230 VAC P 3 Pump 60 Hz 18.7 30 L15-30P
208-230 VAC P 5 Pump 60 Hz 22.3 35 L15-30P
208-230 VAC T 5 Pump 60 Hz 17.3 30 L15-30P
400 VAC P 2 Pump 50 Hz 10.9 20 IEC309
400 VAC P 3 Pump 50 Hz 9.6 15 IEC309
400 VAC P 5 Pump 50 Hz 11.8 15 IEC309
400 VAC T 5 Pump 50 Hz 8.7 15 IEC309
ThermoFlex7500/10000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
200 VAC P 2 Pump 50 Hz 16.2 30 L15-20P
200 VAC P 3 Pump 50 Hz 18.4 30 L15-20P
200 VAC P 5 Pump 50 Hz 22.0 35 L15-30P
200 VAC T 5 Pump 50 Hz 17.0 30 L15-30P
208-230 VAC P 2 Pump 60 Hz 16.2 30 L15-20P
208-230 VAC P 3 Pump 60 Hz 18.4 30 L15-30P
208-230 VAC P 5 Pump 60 Hz 22.0 35 L15-30P
208-230 VAC T 5 Pump 60 Hz 17.0 30 L15-30P
400 VAC P 2 Pump 50 Hz 10.6 20 IEC309
400 VAC P 3 Pump 50 Hz 9.3 15 IEC309
400 VAC P 5 Pump 50 Hz 11.5 20 IEC309
400 VAC T 5 Pump 50 Hz 8.4 15 IEC309
MCA = Minimum Current Ampacity
MOPD = Maximum Overcurrent Protective Device
Values reflect those on the nameplate located on the rear of the chiller.