ThermoFlex-Manual.pdf - 第24页

Section 2 2-6 ThermoFlex Thermo Scientific A: 10000, 60 Hz B: 10000, 50 Hz C: 7500, 60 Hz D: 7500, 50 Hz A B C D 10 20 30 40 °C 50 60 70 80 90 100 °F T emperature Setpoint Cooling Capacity BTU/Hr Watts 40000 35000 30000 2…

100%1 / 138
Section 2
ThermoFlex 2-5
Thermo Scientific
• Specications obtained at sea level using water as the recirculating uid, at a 20°C process
setpoint, 25°C ambient condition, at nominal operating voltage, on chillers with P 2 pumps with no
back pressure. Other uids, uid temperatures, ambient temperatures, altitude, operating voltages or
pumps will affect performance. See Section 3.
• Thermo Fisher Scientic reserves the right to change specications without notice.
Cooling
Capacity
10 20 30 40 °C
50 60 70 80 90 100 °F
Temperature Setpoint
Cooling Capacity
BTU/Hr Watts
A: 5000, 60 Hz
B: 5000, 50 Hz
C: 3500, 60 Hz
D: 3500, 50 Hz
A
B
C
D
7000
6000
5000
4000
3000
2000
1000
24000
20000
16000
12000
8000
4000
A: 2500, 60 Hz
B: 2500, 50 Hz
C: 1400, 60 Hz
D: 1400, 50 Hz
E: 900, 60 Hz
F: 900, 50 Hz
A
C
D
E
F
B
10 20 30 40 °C
50 60 70 80 90 100 °F
Temperature Setpoint
Cooling Capacity
BTU/Hr Watts
12000
10000
8000
6000
4000
2000
3500
3000
2500
2000
1500
1000
500
Section 2
2-6 ThermoFlex
Thermo Scientific
A: 10000, 60 Hz
B: 10000, 50 Hz
C: 7500, 60 Hz
D: 7500, 50 Hz
A
B
C
D
10 20 30 40 °C
50 60 70 80 90 100 °F
Temperature Setpoint
Cooling Capacity
BTU/Hr Watts
40000
35000
30000
25000
20000
15000
10000
5000
Operation below
10°C requires
the use of 50/50
EG/water or
50/50 PG/water
12000
10000
8000
6000
4000
2000
• Specications obtained at sea level using water as the recirculating uid, at a 20°C process
setpoint, 25°C ambient condition, at nominal operating voltage, on chillers with P 2 pumps with
no back pressure (P 3 pumps set to 10 gpm for ThermoFlex15000 to 24000). Other uids, uid
temperatures, ambient temperatures, altitude, operating voltages or pumps will affect performance.
See Section 3.
• Chillers require the use of 50/50 EG/water or 50/50 PG/water below 10°C process temperature to
prevent freezing/glazing of the plate exchanger.
• Thermo Fisher Scientic reserves the right to change specications without notice.
Cooling
Capacity
Operation below
10°C requires
the use of 50/50
EG/water or
50/50 PG/water
A: 24000, 60 Hz
B: 24000, 50 Hz
C: 20000, 60 Hz
D: 20000, 50 Hz
E: 15000, 60 Hz
F: 15000, 50 Hz
A
B
C
D
E
F
10 20 30 40 °C
50 60 70 80 90 100 °F
Temperature Setpoint
Cooling Capacity
30000
25000
20000
15000
10000
5000
BTU/Hr Watts
10000
80000
60000
40000
20000
Section 2
ThermoFlex 2-7
Thermo Scientific
Pumping Capacity
Positive Displacement Pump P 1/P 2
• Pump curves are nominal values. Pressure values for centrifugal pumps are differential pressures
between the inlet and the outlet of the chiller.
• Pump performance results were obtained with no restrictions on the return to the system or with
any options installed. For example, utilizing the DI option will result in a 0.5 gpm ow reduction .
• Specications obtained at sea level using water as the recirculating uid, at a 20°C process
setpoint, 25°C ambient condition, at nominal operating voltage. Other uids, uid temperatures,
ambient temperatures, altitude or operating voltages will affect performance. See Section 3.
• Thermo Fisher Scientic reserves the right to change specications without notice.
5 10 15 20 LPM
1 2 3 4 5 GPM
A
B
C
D
100
80
60
40
20
7
6
5
4
3
2
1
Bar PSIG
Pressure
Flow Rate
Pressure
100
80
60
40
20
7
6
5
4
3
2
1
20 40 60 80 100 LPM
5 10 15 20 25 GPM
Flow Rate
Bar PSID
A
B
C
D
A: P 4, 60 Hz
B: P 4, 50 Hz
C: P 3, 60 Hz
D: P 3, 50 Hz
Pumping Capacity
Centrifugal Pump P 3/P 4
A: P 2/MD 2, 60 Hz
B: P 2/MD 2, 50 Hz
C: P 1/MD 1, 60 Hz
D: P 1/MD 1, 50 Hz