Section 3 3-4 ThermoFlex Thermo Scientific Continued on next page. Electrical Ser vice Requirements (Standard c hillers): ThermoFlex3500/5000 V oltage ±10% Frequency Phase Branch Circuit Requirements Line Cord Plug 200 V …

100%1 / 138
Section 3
ThermoFlex 3-3
Thermo Scientific
Electrical
Requirements
The chiller's construction provides protection against the risk of
electrical shock by grounding appropriate metal parts. The protection
will not function unless the power cord is connected to a properly
grounded outlet. It is the user's responsibility to assure a proper
ground connection is provided.
The chiller must be installed in accordance with the National Electrical Code
and the with reference to the information on the chiller's nameplate located
on the rear.
Locate the chiller so it is near, and has easy access to, its disconnecting device.
The user is responsible to ensure that the line cord provided meets local
 
The chiller is intended for use on a dedicated outlet. The ThermoFlex has
an internal circuit protection that is equivalent (approximately) to the branch
circuit rating. This is to protect the ThermoFlex, and is not intended as a
substitute for branch circuit protection.
DANGER
* Refer to Appendix A for country specific ratings. Continued on next page.
Electrical Service Requirements (Standard chillers):
ThermoFlex900 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
100 VAC 50 Hz 15A 5-15P
115 VAC 60 Hz 15A 5-15P
200 VAC 50 Hz 15A 6-15P
208-230 VAC 60 Hz 15A 6-15P
230 VAC 50 Hz *16A
1
, 15A
2
, 13A
3
-
ThermoFlex1400 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
100 VAC 50 Hz 20A 5-20P
115 VAC 60 Hz 20A 5-20P
200 VAC 50 Hz 15A 6-15P
208-230 VAC 60 Hz 15A 6-15P
230 VAC 50 Hz *16A
1
, 15A
2
, 13A
3
-
ThermoFlex2500 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
200 VAC P 1, P 2 Pump 50 Hz 15A 6-15P
208-230 VAC P 1, P 2 Pump 60 Hz 15A 6-15P
200 VAC T 1 Pump 50 Hz 20A 6-20P
208-230 VAC T 1 Pump 60 Hz 20A 6-20P
230 VAC 50 Hz *16A
1
, 15A
2
, 13A
3
-
Section 3
3-4 ThermoFlex
Thermo Scientific
Continued on next page.
Electrical Service Requirements (Standard chillers):
ThermoFlex3500/5000 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
200 VAC P 1, P 2 Pump 50 Hz 15A 6-15P
200 VAC T 1, P 3 Pump 50 Hz 20A 6-20P
200 VAC P 4 Pump 50 Hz 30A 6-30P
208-230 VAC P 1, P 2 Pump 60 Hz 15A 6-15P
208-230 VAC T 1, P 3 Pump 60 Hz 20A 6-20P
208-230 VAC P 4 Pump 60 Hz 30A 6-30P
230 VAC P 1 - P 4 Pump 50 Hz *16A
1
, 15A
2
, 13A
3
-
ThermoFlex7500/10000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
200 VAC P 2 Pump 50 Hz 16.5 30 L15-20P
200 VAC P 3 Pump 50 Hz 18.7 30 L15-20P
200 VAC P 5 Pump 50 Hz 22.3 35 L15-30P
200 VAC T 5 Pump 50 Hz 17.3 30 L15-30P
208-230 VAC P 2 Pump 60 Hz 16.5 30 L15-20P
208-230 VAC P 3 Pump 60 Hz 18.7 30 L15-30P
208-230 VAC P 5 Pump 60 Hz 22.3 35 L15-30P
208-230 VAC T 5 Pump 60 Hz 17.3 30 L15-30P
400 VAC P 2 Pump 50 Hz 10.9 20 IEC309
400 VAC P 3 Pump 50 Hz 9.6 15 IEC309
400 VAC P 5 Pump 50 Hz 11.8 15 IEC309
400 VAC T 5 Pump 50 Hz 8.7 15 IEC309
ThermoFlex7500/10000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
200 VAC P 2 Pump 50 Hz 16.2 30 L15-20P
200 VAC P 3 Pump 50 Hz 18.4 30 L15-20P
200 VAC P 5 Pump 50 Hz 22.0 35 L15-30P
200 VAC T 5 Pump 50 Hz 17.0 30 L15-30P
208-230 VAC P 2 Pump 60 Hz 16.2 30 L15-20P
208-230 VAC P 3 Pump 60 Hz 18.4 30 L15-30P
208-230 VAC P 5 Pump 60 Hz 22.0 35 L15-30P
208-230 VAC T 5 Pump 60 Hz 17.0 30 L15-30P
400 VAC P 2 Pump 50 Hz 10.6 20 IEC309
400 VAC P 3 Pump 50 Hz 9.3 15 IEC309
400 VAC P 5 Pump 50 Hz 11.5 20 IEC309
400 VAC T 5 Pump 50 Hz 8.4 15 IEC309
MCA = Minimum Current Ampacity
MOPD = Maximum Overcurrent Protective Device
Values reflect those on the nameplate located on the rear of the chiller.
Section 3
ThermoFlex 3-5
Thermo Scientific
ThermoFlex15000/20000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
208-230 VAC P 3 Pump 60 Hz 32.2 60 Hard wire
208-230 VAC P 5 Pump 60 Hz 35.8 60 Hard wire
400 VAC P 3 Pump 50 Hz 15.9 30 Hard wire
400 VAC P 5 Pump 50 Hz 18.1 30 Hard wire
ThermoFlex15000/20000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
208-230 VAC P 3 Pump 60 Hz 28.7 50 Hard wire
208-230 VAC P 5 Pump 60 Hz 32.3 60 Hard wire
400 VAC P 3 Pump 50 Hz 14.5 25 Hard wire
400 VAC P 5 Pump 50 Hz 16.7 30 Hard wire
ThermoFlex24000
(Air-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
208-230 VAC P 3 Pump 60 Hz 43.9 70 Hard wire
208-230 VAC P 5 Pump 60 Hz 47.5 80 Hard wire
400 VAC P 3 Pump 50 Hz 20.1 35 Hard wire
400 VAC P 5 Pump 50 Hz 22.3 40 Hard wire
ThermoFlex24000
(Water-cooled)
Voltage ±10% Frequency Phase MCA MOPD Line Cord
Plug
208-230 VAC P 3 Pump 60 Hz 37.1 70 Hard wire
208-230 VAC P 5 Pump 60 Hz 40.7 70 Hard wire
400 VAC P 3 Pump 50 Hz 18.8 35 Hard wire
400 VAC P 5 Pump 50 Hz 21.0 35 Hard wire
For installation information on variable voltage chillers refer to Appendix B.
Refer to the nameplate label located on the rear of the chiller for specic
electrical requirements.
* United States and Japan only. All other plugs are country specific.
ThermoFlex900 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
115 VAC 60 Hz 15A 5-15P*
100 VAC 50/60 Hz 15A 5-15P*
ThermoFlex1400 Voltage ±10% Frequency Phase Branch Circuit
Requirements
Line Cord
Plug
115 VAC 60 Hz 20A -
100 VAC 50/60 Hz 20A -
Electrical Service Requirements (Variable voltage chillers):