753/754 Calibration Manual 4 Specifications General Specifications All specifications apply from +18 ° C to +28 ° C unless stated otherwise. All specifications assume a 5-minute warmup period. Measurement specifications …

Documenting Process Calibrator
Symbols
3
• Connect the common test lead before the live test
lead and remove the live test lead before the common
test lead.
• Use only current probes, test leads, and adapters
supplied with the Product.
• Do not touch the probes to a voltage source when the
test leads are connected to the current terminals.
• Use only cables with correct voltage ratings.
• Do not use test leads if they are damaged. Examine
the test leads for damaged insulation, exposed metal,
or if the wear indicator shows. Check test lead
continuity.
• Examine the case before you use the Product. Look
for cracks or missing plastic. Carefully look at the
insulation around the terminals.
Symbols
Symbols used on the Product and in this manual are shown in Table 1.
Table 1. Symbols
Symbol Meaning Symbol Meaning
J Earth ground ˜
Common (LO) Input
equipotentiality
AC- alternating current
)
Conforms to relevant North
American Safety Standards.
DC- direct current P
Conforms to European Union
directives.
W
Risk of danger. Important
information. See manual.
f
Pressure
X
Hazardous voltage. Risk of
electrical shock.
~
Do not dispose of this product as
unsorted municipal waste. Go to
Fluke’s website for recycling
information.
,
Application around and removal
from HAZARDOUS LIVE
conductors is permitted.
Conforms to relevant Australian
standards.
T Double insulated ® German certifying body.
CAT II
CAT II equipment is designed to protect against transients from energy-consuming
equipment supplied from the fixed installation, such as TVs, PCs, portable tools, and
other household appliances.

753/754
Calibration Manual
4
Specifications
General Specifications
All specifications apply from +18 °C to +28 °C unless stated otherwise.
All specifications assume a 5-minute warmup period.
Measurement specifications are valid only when Damping is turned on. When damping is turned off, or when the
gannunciator is shown, floor specifications are multiplied by 3. Floor specifications are the second part of the
specifications. The measure pressure, temperature, and frequency functions are specified only with damping on.
Specifications are valid to 110 % of range. The following exceptions are valid to 100 % of range: 300 V dc, 300 V ac,
22 mA source and simulate, 15 V dc source, and temperature measure and source.
To achieve the best noise rejection, use battery power.
Size (H x W x L) ..................................................................... Height = 63.35 mm (2.49 inches) x Width =
136.37 mm (5.37 inches) x Length = 244.96 mm
(9.65 inches)
Weight .................................................................................... 1.23 kg (2.71 lb) (Batteries included)
Display ................................................................................... 480 by 272 pixel graphic LCD, 95 x 54 mm
Power ..................................................................................... Internal battery pack: Lithium Ion, 7.2 V dc, 30 Wh
Environmental Specifications
Operating Altitude ................................................................. 3000 m (9842 ft)
Storage Altitude .................................................................... 13000 m (42650 ft)
Operating Temperature ........................................................ -10 to 50 °C
Storage Temperature ........................................................... -20 to 60 °C
Relative Humidity (Maximum, non-condensing) ................ 90 % to 35 °C
75 % to 40 °C
45 % to 50 °C
Standards and Agency Approval Specifications
Protection Class .................................................................... Pollution Degree II IP 52
Double Insulation Creepage and Clearance ....................... Per IEC 61010-1
Installation Category ............................................................ 300 V CAT II
Design Standards and Compliance .................................... EN/IEC 61010-1:2010, CAN/CSA C22.2 No. 61010-1-04,
ANSI/UL 61010-1:2004
EMI, RFI, EMC ........................................................................ EN 61326-1:2006
RF Fields ................................................................................ Accuracy for all functions is not specified in RF fields >3 V/m
Detailed Specifications
Specifications valid after a 5-minute warmup.
Specifications are valid to 110 % of Range with the following exceptions: 300 V dc measure, 300 V ac measure, 50 kHz
measure and source, 22 mA source and simulate, 15 V dc source, and temperature measure and source which are valid
to 100 % of range.
DC mV Measurement
Range Resolution
% of Reading + Floor
1-Year 2 Year
±100.000 mV
0.001 mV 0.02 % + 0.005 mV 0.03 % + 0.005 mV
Input Impedance: >5 MΩ
Maximum Input Voltage: 300 V, IEC 61010 300 V CAT II
Temperature coefficient: (0.001 % of reading + 0.001% of range) / °C (<18 °C or >28 °C)
Normal mode rejection: >100 dB at 50 or 60 Hz nominal

Documenting Process Calibrator
Detailed Specifications
5
DC Voltage Measurement
Range Resolution
% of Reading + Floor
1-Year 2 Year
±3.00000 V
0.00001 V 0.02 % + 0.00005 V 0.03 % + 0.00005 V
±30.0000 V
0.0001 V 0.02 % + 0.0005 V 0.03 % + 0.0005 V
±300.00 V
0.01 V 0.05 % + 0.05 V 0.07 % + 0.05 V
Input Impedance: >4 MΩ
Maximum Input Voltage: 300 V, IEC 61010 300V CAT II
Temperature coefficient: (0.001 % of reading + 0.0002 % of range) / °C (<18 °C or >28 °C)
Normal mode rejection: >100 dB at 50 or 60 Hz nominal
AC Voltage Measurement
Range
40 Hz – 500 Hz
Resolution
% of Reading + Floor
1-Year 2 Year
3.000 V 0.001 V 0.5 % + 0.002 V 1.0 % + 0.004 V
30.00 V 0.01 V 0.5 % + 0.02 V 1.0 % + 0.04 V
300.0 V 0.1 V 0.5 % + 0.2 V 1.0 % + 0.2 V
Input Impedance: >4 M Ω and <100 pF
Input Coupling: AC
Maximum Input Voltage: 300 V, IEC 61010 300V CAT II
Temperature coefficient: 5 % of specified accuracy / °C (<18 °C or >28 °C)
Specifications apply for 9 % to 100 % of voltage range.
DC Current Measurement
Range Resolution
% of Reading + Floor
1-Year 2 Year
±30.000 mA 1 μA
0.01 % + 5 μA 0.015 % + 7 μA
±100.00 mA 10 μA
0.01 % + 20 μA 0.015 % + 30 μA
Maximum Input: 110 mA
Maximum Burden Voltage: 420 mV at 22 mA
Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
No Fuse
Normal mode rejection: 90 dB at 50 or 60 Hz nominal, and 60 dB at 1200 Hz and 2200 Hz nominal (HART signals)
Resistance Measurement
Range Resolution
% of Reading + Floor
Source
Current
1-Year 2 Year
10.000 Ω 0.001 Ω 0.05 % + 0.050 Ω 0.07 % + 0.070 Ω
3 mA
100.00 Ω 0.01 Ω 0.05 % + 0.05 Ω 0.07 % + 0.07 Ω
1 mA
1.0000 kΩ 0.1 Ω 0.05 % + 0.0005 kΩ 0.07 % + 0.0007 kΩ 500 μA
10.000 kΩ 1 Ω 0.10 % + 0.010 kΩ 0.15 % + 0.015 kΩ 50 μA
Open-circuit voltage: 5 V nominal
Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
Continuity Testing
Tone Resistance
Continuous tone
May or may not get tone
No tone
<25 Ω
25 to 400 Ω
>400 Ω